Note
: The following selections must be made to use the Software Assisted Trigger feature.
x
Edge
trigger type.
x
Normal
trigger mode.
Software Assisted Trigger Setup
Make the following settings from the
Software Assisted Trigger
tab.
1. Touch
the
Enable
checkbox to start software assisted triggering.
2.
Create a trigger window by entering a
Hysteresis
value. This value sets a boundary above and below the
trigger level set in the main Trigger dialog to exclude noise.
3. The
Auto
and
Normal
buttons determine the behavior when trigger crossings are not found in the trigger
source waveform.
x
Auto
mode allows all waveforms through the channel.
x
Normal
mode allows waveforms only with a trigger crossing within the horizontal gate region
through the channel.
4. Set
Start
and
Stop
time values on the
Horizontal Gate
part of the Software Assisted Trigger tab. These
values control where in the waveform the software-assisted trigger processing searches for trigger
crossings.
Example
The following waveform figure has software-assisted triggering disabled:
Summary of Contents for DDA 7 Zi series
Page 1: ...Operator s Manual WavePro SDA and DDA 7 Zi Series Oscilloscopes ...
Page 2: ... L R R H HUD RU D D ...
Page 41: ...Operator s Manual WP700Zi OM E RevA 40 The detachable WavePro Zi front panel ...
Page 376: ...WavePro 7Zi 375 WP700Zi OM E RevA Absolute Offset Relative ...
Page 439: ...Operator s Manual WP700Zi OM E RevA 438 ...
Page 440: ...WavePro 7Zi 439 WP700Zi OM E RevA ...
Page 544: ...Thank you for purchasing a WavePro SDA or DDA 7 Zi Oscilloscope ...