Operator's Manual
WP700Zi-OM-E-RevA
26
Figure 3-2. Setting up Sequence Mode (Adjacent)
1. Touch
Timebase Horizontal Setup
on the menu bar.
2. Under
Sampling Mode
, touch the
Sequence
mode button
.
3. Click
the
Sequence
tab.
4. Under
Acquisition Settings
, touch inside the
Num Segments
data entry field and enter the number of
segments you want to display, using the slider bar at the bottom of the window. Click the keypad button
on the slider bar to enter a value using the pop-up numeric keypad.
Note:
The number of segments you choose to display (80 maximum) can be less than the total
number of segments in the waveform. For example, in the pop-up images above, the number of
display segments is 10, but the total number of segments entered in the timebase dialog's
Num
Segments
field is 100.
5. Touch
the
Enable Timeout
checkbox.
6. Touch
inside
the
Timeout
data entry field and enter a timeout value, using the slider bar at the bottom of
the window. Click the keypad button
on the slider bar to enter a value using the pop-up numeric
keypad.
Note:
Use the sequence mode timeout to automatically interrupt the sequence acquisition if the timeout value is
exceeded without a valid trigger. The timeout period accounts for instances when a
Num Segments
miscount
occurs for some reason and the scope waits indefinitely for an unforthcoming segment. During that time, no
oscilloscope functions are accessible. By means of a timeout value, however, the acquisition will be completed,
the waveform displayed, and control of the oscilloscope returned to the user after the timeout has elapsed.
5. Under
Display Settings
, touch inside the
Display mode
field, and select a sequence mode display from
the pop-up menu.
6.
Touch the S
INGLE
trigger front panel button.
Summary of Contents for DDA 7 Zi series
Page 1: ...Operator s Manual WavePro SDA and DDA 7 Zi Series Oscilloscopes ...
Page 2: ... L R R H HUD RU D D ...
Page 41: ...Operator s Manual WP700Zi OM E RevA 40 The detachable WavePro Zi front panel ...
Page 376: ...WavePro 7Zi 375 WP700Zi OM E RevA Absolute Offset Relative ...
Page 439: ...Operator s Manual WP700Zi OM E RevA 438 ...
Page 440: ...WavePro 7Zi 439 WP700Zi OM E RevA ...
Page 544: ...Thank you for purchasing a WavePro SDA or DDA 7 Zi Oscilloscope ...