Operator's Manual
WP700Zi-OM-E-RevA
474
Local Feature Parameters
lbase Local
Base
Definition
The value of the baseline for a local feature.
Description
The average value of the local baselines for all local features between the parameter cursors
is displayed as lbase. For histograms, each individual baseline value for all local features
between the parameter cursors is provided.
Parameter
Settings
Selecting this parameter from the
Measure
menus prompts you for a hysteresis setting, which
allows you to set the hysteresis value to a specified number of vertical divisions.
Example
lbsep
Local Baseline Separation
Definition
The value of the baseline separation for a local feature.
Description
The average value of the separation of the two baselines used to calculate a local baseline is
displayed for all local features between the parameter cursors. For histograms, each individual
baseline separation value for all local features between the parameter cursors is provided.
Parameter
Settings
Selecting this parameter from the
Measure
menus prompts you for a hysteresis setting, which
allows you to set the hysteresis value to a specified number of vertical divisions.
Example
lmax Local
Maximum
Definition
The maximum value of a local feature.
Description
The maximum value of all local features between the parameter cursors is determined, and
the average value is displayed as lmax. For histograms, the maximum value of each local
feature between the parameter cursors is provided.
Summary of Contents for DDA 7 Zi series
Page 1: ...Operator s Manual WavePro SDA and DDA 7 Zi Series Oscilloscopes ...
Page 2: ... L R R H HUD RU D D ...
Page 41: ...Operator s Manual WP700Zi OM E RevA 40 The detachable WavePro Zi front panel ...
Page 376: ...WavePro 7Zi 375 WP700Zi OM E RevA Absolute Offset Relative ...
Page 439: ...Operator s Manual WP700Zi OM E RevA 438 ...
Page 440: ...WavePro 7Zi 439 WP700Zi OM E RevA ...
Page 544: ...Thank you for purchasing a WavePro SDA or DDA 7 Zi Oscilloscope ...