72
2.5 Subscan Unit
REGIUS MODEL 190 Service Manual Ver.1.00 2004.11.01
•
Remarks for installation
• When reinstalling the sensor assy into the unit, while
pressing the detect roller of the V-sync sensor by
fi
nger as
shown in the right
fi
gure, slide the assy up. If the slide
roller is in upright position, it interferes with the activator
receptor plate, and cannot be slid in.
• When sliding the sensor assy up, align the assy to the left
end as close as possible.
• After con
fi
rming that the top edge of the sensor assy
fi
ts
into the sensor holder (leaf spring) of the main unit, tighten
the
fi
xing screw.
Detect Roller of V-sync Sensor
Sensor Assy Holder
Sensor Assy
Fixing Screw