2.7 Electrical Unit
REGIUS MODEL 190 Service Manual Ver.1.00 2004.11.01
2.7.3
Replacement of SCB2 (System Control Board)
When replacing the CF card at the same time, restore the
backed up data from the REGIUS Console onto the CF card by
following the "Installation/ Service Manual" for REGIUS Con-
sole.
1.
Detach the left cover. (
"2.2.2 Removal of the Left & Right
Covers, p.29"
)
2.
Loosen 6
fi
xing screws, and detach the board cover.
3.
Wear an earth strap.
4.
Unplug three connectors from the SCB2.
•
CN004 CN017, CN021
Fixing Screw 6pcs)
(loosen)
SCB2
CN021
CN004
CN017