111
3.4 Verification of Reading Operation
REGIUS MODEL 190 Service Manual Ver.1.00 2004.11.01
17) Click [DC motors] tab.
18) Click [Close] of "Insertion slot Shutter".
7.
Transports the cassette to the magnetizing (absorb) posi-
tion.
1)
Click [Stepping motors] tab.
2)
Click [Absorb pos.] of "Transport Unit".
3)
Check that "Transport Unit Magnetizing Position ...OK"
is shown on "Information : " window.
8.
Releases the cassette open / close lock, and separates the
front plate from the back plate.
1)
Click [DC motors] tab.
2)
Click [Lock] of "Lock-release motor".
3)
Check that "Lock ...OK" is shown on "Information : "
window.
4)
Click [Stepping motors] tab.
5)
Only when the cassette is mammo type, click [3mm
down] of "Sub scanning".
6)
Click [Escape pos.] of "Transport Unit".
7)
Check that "Transport Unit Escape pos. ...OK" is shown
on "Information : " window.
Transport Motor (PM3)
Feed Escape
Pos. Sen. (S11)
Lock-Release Motor (M4)
Front Plate
Back Plate +
Magnetizing(Absorption)Plate
Summary of Contents for REGIUS 190
Page 7: ...VI REGIUS MODEL 190 Service Manual Ver 1 00 2004 11 01...
Page 13: ...Blank...
Page 27: ...xx REGIUS MODEL 190 Service Manual Ver 1 00 2004 11 01...
Page 28: ...Chap 1 Overview...
Page 29: ...Blank Page...
Page 43: ...16 1 3 Layout of Controlling Components REGIUS MODEL 190 Service Manual Ver 1 00 2004 11 01...
Page 45: ...Blank Page...
Page 52: ...Chap 2 Disassembly and Reassembly...
Page 53: ...Blank Page...
Page 118: ...Chap 3 Maintenance Adjustment...
Page 119: ......
Page 154: ...Chap 4 Troubleshooting...
Page 155: ......
Page 194: ...A Appendix...
Page 195: ......
Page 220: ...0676 55030 B 0412TE...