69
2.5 Subscan Unit
REGIUS MODEL 190 Service Manual Ver.1.00 2004.11.01
9.
Secure the steel belt and the paper onto the top cover so
that they will not fall off.
10.
Loosen a set screw on the coupling.
11.
Remove 6
fi
xing screws, and remove the steel belt deceler-
ating unit.
12.
Carry out reassembly in the reverse order to that described
above.
I m p o r t a n t
When securing the steel belt to the bal-
ance weight, check that the steel belt is
not pinched between the bearing and
washer.
•
Adjustment after installation
Carry out an solid image exposure, and check that there is no
horizontal stripes or artifacts caused by malfunction of the sub-
scan unit.
Coupling
Set Screw
Fixing Screw (6pcs)
Steel Belt Deacceleration Unit
Fixing Screw
Steel Belt
Bearing