114
3.4 Verification of Reading Operation
REGIUS MODEL 190 Service Manual Ver.1.00 2004.11.01
1)
Click [Standby pos.] of "Justify".
2)
Check that "Justi
fi
cation Standby Position ...OK" is
shown on "Information : " window.
3)
Click [Eject pos.] of "Cassette receive Unit".
4)
Check that "Cassette receive Unit Standby Position
...OK" is shown on "Information : " window.
5)
Check that the leading edge of the cassette is located
at the ejection slot.
13.
Ejects the cassette, and returns the transport unit to the
insertion position.
1)
Click [DC motors] tab.
2)
Click [Forward] of "Ejection roller".
3)
Ejects the cassette, and closes the ejection slot shutter.
4)
Click [Stop] of "Ejection roller".
Eject Motor (M3)
Back Plate Sen. (S21)
Eject Shutter Close Sen. (S20)
Summary of Contents for REGIUS 190
Page 7: ...VI REGIUS MODEL 190 Service Manual Ver 1 00 2004 11 01...
Page 13: ...Blank...
Page 27: ...xx REGIUS MODEL 190 Service Manual Ver 1 00 2004 11 01...
Page 28: ...Chap 1 Overview...
Page 29: ...Blank Page...
Page 43: ...16 1 3 Layout of Controlling Components REGIUS MODEL 190 Service Manual Ver 1 00 2004 11 01...
Page 45: ...Blank Page...
Page 52: ...Chap 2 Disassembly and Reassembly...
Page 53: ...Blank Page...
Page 118: ...Chap 3 Maintenance Adjustment...
Page 119: ......
Page 154: ...Chap 4 Troubleshooting...
Page 155: ......
Page 194: ...A Appendix...
Page 195: ......
Page 220: ...0676 55030 B 0412TE...