Cautions Regarding Transpor-
tation and Storage
••••
Temperature and humidity levels must be maintained within the following
specified ranges.
••••
Do not expose the plate to X or
γ
rays.
••••
If the cassette is transported or stored in its packing, ensure that the cassette
is kept in horizontal position. Storage of cassettes in vertical position for
extended periods or in a high-temperature environment may result in warping
of the back plate.
••••
Once the plate is unpacked, store it in the cassette.
••••
Do not store the front and back plates separately. In particular, storage of the
back plate separately in angled position may result in warping.
••••
Be careful not to drop or bump cassettes.
••••
As for previous cassettes, cassettes in use should be stored vertically in a
rack.
Cautions Before Inserting the
Cassette
In order to avoid problems with cassette conveying operation inside the
REGIUS MODEL 190, the following cautions should be observed:
••••
Ensure that the locking mechanism is functioning properly and check for
bending or other damage. Do not insert damaged cassettes into the REGIUS
MODEL 190.
••••
Make sure that there is no paper clipped into the memo clip on the back
plate.
••••
Make sure that there are no foreign objects such as clips between the front
and back plates.
••••
In order to ensure that the back plate properly adheres to the
••••
absorption plate, check that there is nothing (e.g. tape) adhered to the back
plate.
••••
Ensure that the bar code has not become soiled.
Cautions Regarding Disposal
Disposal must be carried out by following the local regulations in each
country.
••••
If the protective layer on the plate is damaged and the photo-stimulated fluo-
rescent coating leaking, be careful to handle the plate so that none of the
photo-stimulated fluorescent coating comes into contact with the body.
••••
Lead foil is incorporated into all plates (RP-3S) except plates for mammo-
grams (RP-3M). Disposal of such plates must be carried out in accordance
with the appropriate waste disposal regulations.
WARNING
When disposing the REGIUS Plate, accept the regulations of each
local government.
Before unsealing the
plate
After unsealing the plate
Temperature
10
:
~40
:
10
:
~30
:
Humidity
Humidity 80%RH or lower (without condensation)
Summary of Contents for REGIUS 190
Page 7: ...VI REGIUS MODEL 190 Service Manual Ver 1 00 2004 11 01...
Page 13: ...Blank...
Page 27: ...xx REGIUS MODEL 190 Service Manual Ver 1 00 2004 11 01...
Page 28: ...Chap 1 Overview...
Page 29: ...Blank Page...
Page 43: ...16 1 3 Layout of Controlling Components REGIUS MODEL 190 Service Manual Ver 1 00 2004 11 01...
Page 45: ...Blank Page...
Page 52: ...Chap 2 Disassembly and Reassembly...
Page 53: ...Blank Page...
Page 118: ...Chap 3 Maintenance Adjustment...
Page 119: ......
Page 154: ...Chap 4 Troubleshooting...
Page 155: ......
Page 194: ...A Appendix...
Page 195: ......
Page 220: ...0676 55030 B 0412TE...