131
4.1 Troubleshooting for Electrical Control & Communication
REGIUS MODEL 190 Service Manual Ver.1.00 2004.11.01
Case 4: Operation status where the network cable has been disconnected during operation.
12) A Reader and REGIUS
Console share the same
host name.
•
Even if the REGIUS Console's host name is set for the Reader, normal read-
ing will be possible.
•
Even if the Reader's host name is set for the REGIUS Console, normal read-
ing will be possible.
13) A Reader and JM share
the same host name.
Regardless of the setting location, normal reading will be possible.
Cause
Remedy
Cause
Remedy
1) Connection with the JM was interrupted
when in the Ready condition.
•
If no operation was being performed, no Reader problems
will result. If the REGIUS Console is not capable of registra-
tion operation, a restart must be carried out.
•
If an error is displayed on the Reader during operation, it
will be necessary to restart the Reader and the REGIUS
Console.
2) Connection with the JM was lost immedi-
ately after the last cassette registration.
If a recorded cassette was read during disconnection
•
The Reader will not perform reading, operation will stop at the
insertion slot, and a "04214" error will occur. Although a restart will
be necessary for all devices, this can be done for the Reader by
simply turning the power Off and then On. A standard restart will
be required for REGIUS Console and JM.
If after disconnection, a non-operating JM is restarted to return
to the reading condition
•
The reading of registered images will be possible; however, it will
not be possible to register a new cassette for the REGIUS Con-
sole, and therefore, a restart must subsequently be carried out.
3) Connection with the JM was lost in Erase
mode while no cassette was inserted
The system will return to Ready mode after the end of Erase
mode. If disconnection continues following this, operation will
be the same as for the above-mentioned disconnection in the
Ready condition.
4) If the network is disconnected for
fi
ve min-
utes after cassette registration
Normal reading will be possible even after
fi
ve minutes.
5) If the network is disconnected for
fi
ve min-
utes after manual registration of cassette.
Normal reading will be possible even after
fi
ve minutes.
6) If the network is disconnected for
fi
ve min-
utes in the Ready status.
Normal reading will be possible even after
fi
ve minutes.
7) If the network is disconnected for
fi
ve min-
utes during shutdown.
It will be possible to turn the power off regardless of when the
disconnection takes place during a POWER OFF operation.
8) During initialization after POWERON, the
network is disconnected for
fi
ve minutes
from when there is one "
■
". (Operation is
the same for up to ten "
■
".)
A 24000 error occurs for the Reader, and if the network cable is
subsequently connected to re-establish connection with the
JM, the Ready condition will be achieved and normal operation
will be possible.
Summary of Contents for REGIUS 190
Page 7: ...VI REGIUS MODEL 190 Service Manual Ver 1 00 2004 11 01...
Page 13: ...Blank...
Page 27: ...xx REGIUS MODEL 190 Service Manual Ver 1 00 2004 11 01...
Page 28: ...Chap 1 Overview...
Page 29: ...Blank Page...
Page 43: ...16 1 3 Layout of Controlling Components REGIUS MODEL 190 Service Manual Ver 1 00 2004 11 01...
Page 45: ...Blank Page...
Page 52: ...Chap 2 Disassembly and Reassembly...
Page 53: ...Blank Page...
Page 118: ...Chap 3 Maintenance Adjustment...
Page 119: ......
Page 154: ...Chap 4 Troubleshooting...
Page 155: ......
Page 194: ...A Appendix...
Page 195: ......
Page 220: ...0676 55030 B 0412TE...