36
2.3 Insertion / Ejection Unit
REGIUS MODEL 190 Service Manual Ver.1.00 2004.11.01
•
Controlling parts comprising the ejection assy
Actuator
Eject Rollers
Eject Shutter
1
2
3
Part Name
Function
Access to the Part
1 Eject Shutter Close
sensor (S20)
Detects the eject
shutter close posi-
tion.
1) Remove the insertion / ejection unit.
•Refer to
"2.3.2 Removal of the Insertion/Ejection Unit, p.37"
.
2) Remove the guide plate for eject sensor wiring.
•Refer to
"2.3.8 Replacement of Ejection Motor (M3), p.45"
3) Remove the eject sensor bracket.
•Refer to
"2.3.8 Replacement of Ejection Motor (M3), p.45"
.
4) Remove the sensor from the eject sensor bracket.
5) Unplug a connector.
•JP54
2 Back Plate Detect Sen-
sor (S21)
Detects presence
of the back plate
through actuator.
1) Remove the insertion / ejection unit.
•Refer to
"2.3.2 Removal of the Insertion/Ejection Unit, p.37"
.
2) Remove the actuator bracket.
3) Remove the sensor from the actuator bracket.
4) Unplug a connector.
•JP51
3 Eject Motor (M3)
Drives the eject
rollers.
Refer to
"2.3.8 Replacement of Ejection Motor (M3), p.45"
Summary of Contents for REGIUS 190
Page 7: ...VI REGIUS MODEL 190 Service Manual Ver 1 00 2004 11 01...
Page 13: ...Blank...
Page 27: ...xx REGIUS MODEL 190 Service Manual Ver 1 00 2004 11 01...
Page 28: ...Chap 1 Overview...
Page 29: ...Blank Page...
Page 43: ...16 1 3 Layout of Controlling Components REGIUS MODEL 190 Service Manual Ver 1 00 2004 11 01...
Page 45: ...Blank Page...
Page 52: ...Chap 2 Disassembly and Reassembly...
Page 53: ...Blank Page...
Page 118: ...Chap 3 Maintenance Adjustment...
Page 119: ......
Page 154: ...Chap 4 Troubleshooting...
Page 155: ......
Page 194: ...A Appendix...
Page 195: ......
Page 220: ...0676 55030 B 0412TE...