165
4.4 Error Code
REGIUS MODEL 190 Service Manual Ver.1.00 2004.11.01
14010
Signal process error
QRV Over
fl
ow
The sensitivity has been mistakenly
corrected. Check the following.
1) Con
fi
rm that the ADB2's high volt-
age On/Off switch is On (check by
reading the solid density).
2) Con
fi
rm the absence of problems
in the exposure X-ray.
3) Replace the plate and con
fi
rm
operation.
4) Check the connectors in the vicin-
ity of the PMT and ADB2.
14011
Signal process error
QRV Under
fl
ow
The sensitivity has been mistakenly
corrected. Check the following.
1) Con
fi
rm that the ADB2's high volt-
age On/Off switch is On (check by
reading the solid density).
2) Con
fi
rm the absence of problems
in the exposure X-ray.
3) Replace the plate and con
fi
rm
operation.
4) Check the connectors in the vicin-
ity of the PMT and ADB2.
Code
Medium Classifi-
cation
Detailed Clas-
sification
Countermeasures, etc.
Summary of Contents for REGIUS 190
Page 7: ...VI REGIUS MODEL 190 Service Manual Ver 1 00 2004 11 01...
Page 13: ...Blank...
Page 27: ...xx REGIUS MODEL 190 Service Manual Ver 1 00 2004 11 01...
Page 28: ...Chap 1 Overview...
Page 29: ...Blank Page...
Page 43: ...16 1 3 Layout of Controlling Components REGIUS MODEL 190 Service Manual Ver 1 00 2004 11 01...
Page 45: ...Blank Page...
Page 52: ...Chap 2 Disassembly and Reassembly...
Page 53: ...Blank Page...
Page 118: ...Chap 3 Maintenance Adjustment...
Page 119: ......
Page 154: ...Chap 4 Troubleshooting...
Page 155: ......
Page 194: ...A Appendix...
Page 195: ......
Page 220: ...0676 55030 B 0412TE...