70
2.5 Subscan Unit
REGIUS MODEL 190 Service Manual Ver.1.00 2004.11.01
2.5.4
Replacement of Peel Detect Sensor 1/2 (S13 /S14)
Two peel detect sensor on the right and left are symmetric.The
one on the left is a peel detect sensor-L (S13), and the one on
the right is a peel detect sensor-R (S14).
1.
Remove the left or right cover. (
"2.2.2 Removal of the Left &
Right Covers, p.29"
)
2.
Remove a
fi
xing screw "a", detach the peel detect sensor-L
mount plate or peel detect sensor-R mount plate.
3.
Remove a
fi
xing screw "b" which is securing the peel detect
sensor, and detach the peel detect sensor.
4.
Remove the wire bundle from the wire clamp.
5.
Unplug the connector from the peel detect sensor.
•
JP41 (S13) or JP42 (S14)
6.
Carry out reassembly in the reverse order to that described
above.
•
Adjustment after installation
Always carry out
"3.6 Adjusting the Peel-Detection Rollers,
p.117"
.
Peel Detect Sensor
Fixing Screw "a"
Fixing Screw "b"
Peel Detect Sensor Mout Plate
Summary of Contents for REGIUS 190
Page 7: ...VI REGIUS MODEL 190 Service Manual Ver 1 00 2004 11 01...
Page 13: ...Blank...
Page 27: ...xx REGIUS MODEL 190 Service Manual Ver 1 00 2004 11 01...
Page 28: ...Chap 1 Overview...
Page 29: ...Blank Page...
Page 43: ...16 1 3 Layout of Controlling Components REGIUS MODEL 190 Service Manual Ver 1 00 2004 11 01...
Page 45: ...Blank Page...
Page 52: ...Chap 2 Disassembly and Reassembly...
Page 53: ...Blank Page...
Page 118: ...Chap 3 Maintenance Adjustment...
Page 119: ......
Page 154: ...Chap 4 Troubleshooting...
Page 155: ......
Page 194: ...A Appendix...
Page 195: ......
Page 220: ...0676 55030 B 0412TE...