156
4.4 Error Code
REGIUS MODEL 190 Service Manual Ver.1.00 2004.11.01
4.4.4
Mechanical Control Error Code (04150 ~ 04165)
04150 ~ 04165 indicates an error occurred on the
fi
rmware.
Error
code
Error name
Error description
Action
04150
Lock initializing pro-
cess
Initialization of the lock motor
cannot be carried out as a result
of a problem in the
fi
rmware.
Open the front door, remove any cas-
sette that may be present inside the
equipment, and then carry out a restart.
04151
Insertion shutter initial-
izing process
Initialization of the insertion slot
shutter motor cannot be carried
out as a result of a problem in the
fi
rmware.
Open the front door, remove any cas-
sette that may be present inside the
equipment, and then carry out a restart.
04152
Cassette receive Unit
initializing process
Initialization of the elevation-bed
motor cannot be carried out as a
result of a problem in the
fi
rm-
ware.
Open the front door, remove any cas-
sette that may be present inside the
equipment, and then carry out a restart.
04153
Justi
fi
cation initializing
process
Initialization of the justi
fi
cation
motor cannot be carried out as a
result of a problem in the
fi
rm-
ware.
Open the front door, remove any cas-
sette that may be present inside the
equipment, and then carry out a restart.
04154
Transport Unit initializ-
ing process
Initialization of the Transportation
Motor cannot be carried out as a
result of a problem in the
fi
rm-
ware.
Open the front door, remove any cas-
sette that may be present inside the
equipment, and then carry out a restart.
04155
Subscan initializing
process
Initialization of the auxiliary scan-
ning motor cannot be carried out
as a result of a problem in the
fi
rmware.
Open the front door, remove any cas-
sette that may be present inside the
equipment, and then carry out a restart.
04156
Feed in operation 1
start NG
Draw-in operation 1 could not be
carried out by the insertion slot
motor as a result of a problem in
the
fi
rmware.
Open the front door, remove any cas-
sette that may be present inside the
equipment, and then carry out a restart.
04157
Feed in operation 2
start NG
Draw-in operation 2 could not be
carried out by the insertion slot
motor as a result of a problem in
the
fi
rmware.
Open the front door, remove any cas-
sette that may be present inside the
equipment, and then carry out a restart.
04158
Insertion slot ejection
start NG
An eject operation could not be
carried out by the insertion slot
motor as a result of a problem in
the
fi
rmware.
Open the front door, remove any cas-
sette that may be present inside the
equipment, and then carry out a restart.
04159
Ejection start NG
An eject operation could not be
carried out by the removal-port
motor as a result of a problem in
the
fi
rmware.
Open the front door, remove any cas-
sette that may be present inside the
equipment, and then carry out a restart.
04160
Insertion slot shutter
open NG
An opening operation could not
be carried out by the insertion slot
motor as a result of a problem in
the
fi
rmware.
Open the front door, remove any cas-
sette that may be present inside the
equipment, and then carry out a restart.
04161
Insertion slot shutter
close NG
A closing operation could not be
carried out by the insertion slot
shutter motor as a result of a
problem in the
fi
rmware.
Open the front door, remove any cas-
sette that may be present inside the
equipment, and then carry out a restart.
Summary of Contents for REGIUS 190
Page 7: ...VI REGIUS MODEL 190 Service Manual Ver 1 00 2004 11 01...
Page 13: ...Blank...
Page 27: ...xx REGIUS MODEL 190 Service Manual Ver 1 00 2004 11 01...
Page 28: ...Chap 1 Overview...
Page 29: ...Blank Page...
Page 43: ...16 1 3 Layout of Controlling Components REGIUS MODEL 190 Service Manual Ver 1 00 2004 11 01...
Page 45: ...Blank Page...
Page 52: ...Chap 2 Disassembly and Reassembly...
Page 53: ...Blank Page...
Page 118: ...Chap 3 Maintenance Adjustment...
Page 119: ......
Page 154: ...Chap 4 Troubleshooting...
Page 155: ......
Page 194: ...A Appendix...
Page 195: ......
Page 220: ...0676 55030 B 0412TE...