97
3.1 Maintenance Items
REGIUS MODEL 190 Service Manual Ver.1.00 2004.11.01
3.1.5
Cleaning of Power Supply Fan Filter
Clean the
fi
lter once a year following the procedure below.
1.
Remove the rear cover. (refer to p.28)
2.
Remove one
fi
xing screw "a", and loosen 2
fi
xing screws
"b" to remove the fan bracket.
•
Be careful not to apply excessive force to the DC power
supply cooling fan connector that is located on the back
of the fan bracket.
3.
Unplug one relay connector of the DC power supply cooling
fan.
•
JJ60
4.
Remove 2
fi
xing screws retaining the power supply cooling
fan, and remove the power supply cooling fan
fi
lter.
5.
Vacuum the dust accumulated on the back side (fan side)
of the
fi
lter using a vacuum cleaner.
6.
Install the power supply cooling fan
fi
lter at its original posi-
tion, and install the fan bracket in the reverse order of
removal.
Fixing Screw "a"
(remove)
Fixing Screw "b"
(loosen)
Relay Connector
Fan Bracket
Fixing Screw
Power Supply
Cooling Fan
Summary of Contents for REGIUS 190
Page 7: ...VI REGIUS MODEL 190 Service Manual Ver 1 00 2004 11 01...
Page 13: ...Blank...
Page 27: ...xx REGIUS MODEL 190 Service Manual Ver 1 00 2004 11 01...
Page 28: ...Chap 1 Overview...
Page 29: ...Blank Page...
Page 43: ...16 1 3 Layout of Controlling Components REGIUS MODEL 190 Service Manual Ver 1 00 2004 11 01...
Page 45: ...Blank Page...
Page 52: ...Chap 2 Disassembly and Reassembly...
Page 53: ...Blank Page...
Page 118: ...Chap 3 Maintenance Adjustment...
Page 119: ......
Page 154: ...Chap 4 Troubleshooting...
Page 155: ......
Page 194: ...A Appendix...
Page 195: ......
Page 220: ...0676 55030 B 0412TE...