1.1 Principal Specifications
REGIUS MODEL 190 Service Manual Ver.1.00 2004.11.01
•
Main Scan Unit
•
Beam-Reception Unit
•
Erasing Unit
Method
Laser drive using polygon mirror technology
Incident Angle of Laser
Approximately 5º from the plate's normal
One-Line Read
Frequency
2 types; 7.54ms/line (133kHz), 5.20ms/line (192kHz)
f-
θθ
θ
θ
Lens
3 groups of 3 (f = 253.6 mm)
Tilt compensation using cylindrical mirrors
Effective Scan Ratio
82.8% ±49.7º)
No. of Polygon Faces
7 (inscribed circle diameter: 82ø)
Rotation Speed of
Polygon
2 types; 1,137rpm, 1,648rpm
Laser Wavelength
660nm
On-Plate Laser Inten-
sity
35mW or higher
Imaging beam
diameter100
100
∫
or less (for both main and sub scanning)
Method
Condensing Beam guide + Condenser tube + Short photomulti-
plier
Beam Guide
Uses full re
fl
ection from acrylic plates
Condenser Tube
Ammonite condenser with multiple coatings on the inner surface
Excitation-beam elimi-
nation
Realized using a excitation-beam cut
fi
lter
Method
Erasure by halogen lamps
Lamp Spec.
2 x 300W halogen lamps
Filter Spec.
Cuts light in the photo-multiplier's sensitivity range
Summary of Contents for REGIUS 190
Page 7: ...VI REGIUS MODEL 190 Service Manual Ver 1 00 2004 11 01...
Page 13: ...Blank...
Page 27: ...xx REGIUS MODEL 190 Service Manual Ver 1 00 2004 11 01...
Page 28: ...Chap 1 Overview...
Page 29: ...Blank Page...
Page 43: ...16 1 3 Layout of Controlling Components REGIUS MODEL 190 Service Manual Ver 1 00 2004 11 01...
Page 45: ...Blank Page...
Page 52: ...Chap 2 Disassembly and Reassembly...
Page 53: ...Blank Page...
Page 118: ...Chap 3 Maintenance Adjustment...
Page 119: ......
Page 154: ...Chap 4 Troubleshooting...
Page 155: ......
Page 194: ...A Appendix...
Page 195: ......
Page 220: ...0676 55030 B 0412TE...