75
2.6 Optical Unit
REGIUS MODEL 190 Service Manual Ver.1.00 2004.11.01
2.6
Optical Unit
2.6.1
Replacement of Photomultiplier (PMT)
I m p o r t a n t
When handling the PMT, be cautious not
to touch the light collector (glass surface)
with
fi
ngers or any object. Once scratch
is generated, it may result in deteriorated
image quality.
1.
Remove the left cover. (
"2.2.2 Removal of the Left & Right
Covers, p.29"
)
2.
Remove 4
fi
xing screws, and loosen 2
fi
xing screws to
remove the board cover.
3.
Unplug two connectors from the analog/digital board
(ADB2).
•
J1, J3
4.
Remove four
fi
xing screws, and remove the photomultiplier
from the optical unit.
I m p o r t a n t
When placing the photomultiplier on the
table, always place it with its light collec-
tor side (glass surface) facing up.
5.
Carry out reassembly in the reverse order to that described
above.
•
Adjustment after replacement
Always carry out the calibration list below following the "Instal-
lation / Service Manual" or REGIUS Console.
• Uniformity Calibration
• Sensitivity Calibration
Photomultiplier
Connector (J1)
Connector (J3)
Fixing Screws (4pcs)
Summary of Contents for REGIUS 190
Page 7: ...VI REGIUS MODEL 190 Service Manual Ver 1 00 2004 11 01...
Page 13: ...Blank...
Page 27: ...xx REGIUS MODEL 190 Service Manual Ver 1 00 2004 11 01...
Page 28: ...Chap 1 Overview...
Page 29: ...Blank Page...
Page 43: ...16 1 3 Layout of Controlling Components REGIUS MODEL 190 Service Manual Ver 1 00 2004 11 01...
Page 45: ...Blank Page...
Page 52: ...Chap 2 Disassembly and Reassembly...
Page 53: ...Blank Page...
Page 118: ...Chap 3 Maintenance Adjustment...
Page 119: ......
Page 154: ...Chap 4 Troubleshooting...
Page 155: ......
Page 194: ...A Appendix...
Page 195: ......
Page 220: ...0676 55030 B 0412TE...