3.2 Set Up of dummy SBC2 program
REGIUS MODEL 190 Service Manual Ver.1.00 2004.11.01
3.2.2
Connection
Connect the maintenance PC to the REGIUS 190.
1.
Move the REGIUS 190's power breaker to the OFF posi-
tion.
2.
Remove the left side cover. "
2.2.2 Removal of the Left &
Right Covers, p.29"
)
3.
Loosen 5
fi
xing screws, and remove the cover of the
mechanical control box.
4.
Move the SW2-1 dip switch on the top of the MCB2 to the
ON (right) position.
5.
Unplug the connector from CN4 on the MCB2.
•
CN4
6.
Connect the Unit Checker RS-232C cable between CN4
and the EIA-232C port of the maintenance PC
ON
SW2-1
MCB2
Left side
MCB2
Left side
Maintenance PC
EIA-232C
CN4