3.5 Checking the Stand-Alone Operation
REGIUS MODEL 190 Service Manual Ver.1.00 2004.11.01
3.5
Checking the Stand-Alone Operation
Normally, when the REGIUS 190 which is not networked to the
REGIUS Console is started up, a network error 24000 occurs,
and it will not start even the cassette is inserted.
When it is necessary to operate the REGIUS 190 as stand-
alone, follow the procedures described below.
I m p o r t a n t
Do not change the setting of DIP SW
other than SW2- 5 on the MCB2. (for the
rest of DIP SW, refer to "5.2 DIP Switch
Setting", p.184)
I m p o r t a n t
To terminate the dummy SBC2 program,
con
fi
rm that the all dips of DIP SW (SW2)
are set to OFF (left).
1.
Remove the left cover. ("2.2.2 Removal of the Left & Right
Covers", p.29)
2.
Loosen 5
fi
xing screws, and detach the cover for mechani-
cal control box.
3.
Turn the DIP SW (SW2-5) on the MCB2 (Mechanical Con-
trol Board) to ON (right).
4.
Press operation" button of the REGIUS 190.
•
Insert the cassette into the insertion slot. The cassette
will be drawn into the device, and series of operations
including ejection of cassette will be initiated.
ON
SW2-5
MCB2
Left side
Summary of Contents for REGIUS 190
Page 7: ...VI REGIUS MODEL 190 Service Manual Ver 1 00 2004 11 01...
Page 13: ...Blank...
Page 27: ...xx REGIUS MODEL 190 Service Manual Ver 1 00 2004 11 01...
Page 28: ...Chap 1 Overview...
Page 29: ...Blank Page...
Page 43: ...16 1 3 Layout of Controlling Components REGIUS MODEL 190 Service Manual Ver 1 00 2004 11 01...
Page 45: ...Blank Page...
Page 52: ...Chap 2 Disassembly and Reassembly...
Page 53: ...Blank Page...
Page 118: ...Chap 3 Maintenance Adjustment...
Page 119: ......
Page 154: ...Chap 4 Troubleshooting...
Page 155: ......
Page 194: ...A Appendix...
Page 195: ......
Page 220: ...0676 55030 B 0412TE...