151
4.4 Error Code
REGIUS MODEL 190 Service Manual Ver.1.00 2004.11.01
4.4.2
Mechanical Control Error Codes
04001 ~ 04069 indicates an error occurred on the sensors.
Error
code
Error name
Error description
Action
04001
Feed HP sensor NG
A time-out occurred for ON or OFF
detection by the Feed HP sensor
while the transport unit is moving
to its receiving position.
Check MCB2:CN11-, pins 32 through 34
(HP sensor) and MCB2:CN15 (Trans-
portation Motor)
04011
Feed Escape Position
sensors NG (at initial-
ization)
A time-out occurred for ON detec-
tion by the Feed Escape Position
sensor while the transport unit is
moving to escape position.
Check MCB2:CN11-, pins 35 through 37
(Feed Escape Position sensors) and
MCB2:CN15 (Transportation Motor)
04012
Feed Escape Position
sensors NG (while
moving to absorption
position)
1) The Feed Escape Position sen-
sor was already ON at the start
of motion by the transport unit
from the receive position to the
absorption position.
2) A limit-over occurred for ON
detection by the Feed Escape
Position sensor during motion
by the transport unit from the
receive position to the absorp-
tion position.
Check MCB2:CN11-, pins 35 through 37
(Feed Escape Position sensors) and
MCB2:CN15 (Transportation Motor)
04013
Feed Escape Position
sensors NG (while
moving to escape
position)
A limit-over occurred for OFF
detection by the Feed Escape Posi-
tion sensor during motion by the
transport unit from the absorption
position to the escape position.
Check MCB2:CN11-, pins 35 through 37
(Feed Escape Position sensors) and
MCB2:CN15 (Transportation Motor)
04014
Feed Escape Position
sensors NG (while
moving to eject posi-
tion 1)
A limit-over occurred for OFF
detection by the Feed Escape
Position sensor during motion by
the transport unit from the absorp-
tion position to the escape posi-
tion.
Check MCB2:CN11-, pins 35 through 37
(Feed Escape Position sensors) and
MCB2:CN15 (Transportation Motorr)
04015
Feed Escape Position
sensors NG (while
moving to eject posi-
tion 2)
During motion by the transport unit
from the absorption position to the
eject position, ON --> OFF detec-
tion by the Feed Escape Position
sensor occurred within 70 mm
from the absorption position.
Check MCB2:CN11-, pins 35 through 37
(Feed Escape Position sensors) and
MCB2:CN15 (Transportation Motor)
04021
Justi
fi
cation R/L sen-
sor NG (at reading)
1) The Justi
fi
cation Sensor was
OFF after the completion of jus-
ti
fi
cation operation during cas-
sette intake.
2) The justi
fi
cation sensor was
OFF after closing of the inser-
tion slot shutter during cas-
sette take-in.
•
Check MCB2:CN11-, pins 9 through
11 (Justi
fi
cation Sensor- L);
MCB2:CN11-, pins 12 through 14
(Justi
fi
cation Sensor-R); and
MCB2:CN17 (Justi
fi
cation Motor)
•
Check the loose of the drive pulley
Refer to "
2.4.3 Replacement of the
Justi
fi
cation Motor (PM2), p.55"
and
"
3.8 Checking and Adjusting the
Justi
fi
cation Allowance, p.124"
Summary of Contents for REGIUS 190
Page 7: ...VI REGIUS MODEL 190 Service Manual Ver 1 00 2004 11 01...
Page 13: ...Blank...
Page 27: ...xx REGIUS MODEL 190 Service Manual Ver 1 00 2004 11 01...
Page 28: ...Chap 1 Overview...
Page 29: ...Blank Page...
Page 43: ...16 1 3 Layout of Controlling Components REGIUS MODEL 190 Service Manual Ver 1 00 2004 11 01...
Page 45: ...Blank Page...
Page 52: ...Chap 2 Disassembly and Reassembly...
Page 53: ...Blank Page...
Page 118: ...Chap 3 Maintenance Adjustment...
Page 119: ......
Page 154: ...Chap 4 Troubleshooting...
Page 155: ......
Page 194: ...A Appendix...
Page 195: ......
Page 220: ...0676 55030 B 0412TE...