23
1.5 Description of Operation
REGIUS MODEL 190 Service Manual Ver.1.00 2004.11.01
1.5.3
Initializing Sequence
Initializing sequence will be completed by turning ON the power breaker and pressing goperation" button of the
REGIUS Model 190.
Use the dummy SCB2 software to check mechanical operations. Refer to Refer to "3.3 Veri
fi
cation of Initializing
Sequence".
1.5.4
Erase Operation
Erase operation (both ERASE MODE 1 and ERASE MODE 2) follows the same sequence as the normal opera-
tion. Erasing will be completed in step 7 of
"1.5.2 Normal Operation"
.
1
Insertion Slot Shutter Initialization of the insertion slot shutter is implemented.
2
Justi
fi
cation Assy
Justi
fi
cation assy moves to HP.
3
Transport Unit
Transport unit moves to HP.
4
Elevator Stand
Elevator stand moves to HP, and initialization is implemented. After completing ini-
tialization, the elevator stand moves to receive position.
5
Subscanning Unit
Initialization of the subscanning unit is implemented.
6
Transport Unit
Initialization of the transport unit is implemented.
7
Elevator Stand
Elevator stand moves to HP.
8
Lock-Release Motor
Lock-release motor is locked.
Summary of Contents for REGIUS 190
Page 7: ...VI REGIUS MODEL 190 Service Manual Ver 1 00 2004 11 01...
Page 13: ...Blank...
Page 27: ...xx REGIUS MODEL 190 Service Manual Ver 1 00 2004 11 01...
Page 28: ...Chap 1 Overview...
Page 29: ...Blank Page...
Page 43: ...16 1 3 Layout of Controlling Components REGIUS MODEL 190 Service Manual Ver 1 00 2004 11 01...
Page 45: ...Blank Page...
Page 52: ...Chap 2 Disassembly and Reassembly...
Page 53: ...Blank Page...
Page 118: ...Chap 3 Maintenance Adjustment...
Page 119: ......
Page 154: ...Chap 4 Troubleshooting...
Page 155: ......
Page 194: ...A Appendix...
Page 195: ......
Page 220: ...0676 55030 B 0412TE...