7-2
8
PAT-T 4KW
Controlling the constant voltage/current
VOLT:EXT:SOUR
Sets the constant voltage control mode. This cannot be set when the output is on.
Command
[SOURce:]VOLTage:EXTernal:SOURce
{NONE|VOLTage|RESistance|IRESistance}
[SOURce:]VOLTage:EXTernal:SOURce?
Parameter
Val
u
e:
NONE
Panel control (default)
VOLTage
External voltage control
RESistance External resistance control 10 k
Ω
→
MAX OUT
IRESistance External resistance control 10 k
Ω
→
0 OUT (FAIL
SAFE)
For the setting that is applied when *RST is sent, see Table A-16.
Response
Returns the constant voltage control mode as character data in response to
VOLT:EXT:SOUR?.
CURR:EXT:SOUR
Sets the constant current control mode. This cannot be set when the output is on.
Command
[SOURce:]CURRent:EXTernal:SOURce
{NONE|VOLTage|RESistance|IRESistance}
[SOURce:]CURRent:EXTernal:SOURce?
Parameter
Val
u
e:
NONE
Panel control (default)
VOLTage
External voltage control
RESistance External resistance control 10 k
Ω
→
MAX OUT
IRESistance External resistance control 10 k
Ω
→
0 OUT (FAIL
SAFE)
For the setting that is applied when *RST is sent, see Table A-16.
Response
Returns the constant current control mode as character data in response to
CURR:EXT:SOUR?.
Summary of Contents for PAT160-25T
Page 17: ...General Description This chapter describes an overview and describes the features...
Page 22: ...1 6 PAT T 4kW...
Page 34: ...2 12 PAT T 4kW...
Page 106: ...6 12 PAT T 4kW...
Page 158: ...7 52 PAT T 4KW...
Page 159: ...Maintenance This chapter describes maintenance such as cleaning inspecting and calibrating...
Page 171: ...Specifications This chapter describes the specifications and gives the dimensions of the PAT...