PAT-T 4KW
7-37
7
Remot
e
Cont
ro
l
INIT:SEQ2 / INIT:NAME OUTP
Starts the trigger function of the sequence 2 group.
If TRIG:SEQ2:SOUR or TRIG:OUTP:SOUR is set to IMM, the delay action starts
immediately. If set to BUS, the delay action starts after waiting for a software
trigger.
Command
INITiate[:IMMediate]:SEQuence2
INITiate[:IMMediate]:NAME OUTPut
TRIG:SEQ2 / TRIG:OUTP
Executes a software trigger to the sequence 2 group.
Command
TRIGger:SEQuence2[:IMMediate]
TRIGger:OUTPut[:IMMediate]
7.14.3 Measurement (Sequence 3: ACQuire Settings)
The sequence 3 group can control the measurement conditions using triggers.
TRIG:SEQ3:SOUR / TRIG:ACQ:SOUR
Sets the condition (trigger source) for actually starting the measurement after the
sequence 3 group receives INIT:SEQ3 / INIT:NAME ACQ.
Command
TRIGger:SEQuence3:SOURce {IMMediate|BUS}
TRIGger:SEQuence3:SOURce?
TRIGger:ACQuire:SOURce {IMMediate|BUS}
TRIGger:ACQuire:SOURce?
Parameter
Val
u
e:
IMM
Start the measurement immediately (default)
BUS
Wait for a software trigger (*TRG, TRIG:SEQ3,
TRIG:ACQ, or IEEE488.1 get (Group Execute
Trigger)) to start the measurement
For the setting that is applied when *RST is sent, see Table A-16.
Response
Returns the trigger source of the sequence 3 group as character data in response to
TRIG:SEQ3:SOUR? / TRIG:ACQ:SOUR?.
See
Summary of Contents for PAT160-25T
Page 17: ...General Description This chapter describes an overview and describes the features...
Page 22: ...1 6 PAT T 4kW...
Page 34: ...2 12 PAT T 4kW...
Page 106: ...6 12 PAT T 4kW...
Page 158: ...7 52 PAT T 4KW...
Page 159: ...Maintenance This chapter describes maintenance such as cleaning inspecting and calibrating...
Page 171: ...Specifications This chapter describes the specifications and gives the dimensions of the PAT...