PAT-T 4KW
7-25
7
Remot
e
Cont
ro
l
Current setting limit
•
When the phase input mode is changed to the single-phase input mode from the
three-phase input mode, the setting value of the output current and the
overcurrent protection (OCP) is limited to 75 % of the rated value of output
current and the overcurrent protection (OCP) of the three-phase input mode.
CURR:LIM
Queries the current setting limit. The current setting limit is approximately 95 % of
the OCP trip point.
This command is valid when CURR:LIM:AUTO ON is specified.
Command
[SOURce:]CURRent:LIMit[:AMPlitude]?
Response
Returns the current limit value in the NR3 form in response to CURR:LIM?.
Unit
A
CURR:LIM:AUTO
Sets whether to enable the current setting limit. If enabled, the current setting limit
is automatically set to approximately 95 % of the OCP trip point. This cannot be set
when the output is on.
Command
[SOURce:]CURRent:LIMit:AUTO {ON|OFF|1|0}
[SOURce:]CURRent:LIMit:AUTO?
Parameter
Val
u
e:
ON (1)
Limit the current setting
OFF (0)
Not limit the current setting (default)
For the setting that is applied when *RST is sent, see Table A-16.
Response
Returns whether to enable the current setting limit in the NR1 form in response to
CURR:LIM:AUTO?.
Summary of Contents for PAT160-25T
Page 17: ...General Description This chapter describes an overview and describes the features...
Page 22: ...1 6 PAT T 4kW...
Page 34: ...2 12 PAT T 4kW...
Page 106: ...6 12 PAT T 4kW...
Page 158: ...7 52 PAT T 4KW...
Page 159: ...Maintenance This chapter describes maintenance such as cleaning inspecting and calibrating...
Page 171: ...Specifications This chapter describes the specifications and gives the dimensions of the PAT...