7-24
PAT-T 4KW
VOLT:LIM:AUTO
Sets whether to enable the voltage limit setting. If enabled, the voltage limit setting
is automatically set to approximately 95 % of the OVP trip point. This cannot be set
when the output is on.
Command
[SOURce:]VOLTage:LIMit:AUTO {ON|OFF|1|0}
[SOURce:]VOLTage:LIMit:AUTO?
Parameter
Val
u
e:
ON (1)
Limit the voltage setting
OFF (0)
Not limit the voltage setting (default)
For the setting that is applied when *RST is sent, see Table A-16.
Response
Returns whether to enable the voltage limit setting in the NR1 form in response to
VOLT:LIM:AUTO?.
Overcurrent protection settings
CURR:PROT
Sets the overcurrent protection (OCP) value.
Command
[SOURce:]CURRent:PROTection[:LEVel]
{<numeric>|MIN|MAX}
[SOURce:]CURRent:PROTection[:LEVel]? [MIN|MAX]
Parameter
Val
u
e:
10 % to 111.5 % of the rated output current
(The default value is 111.5 % of the rated output current.)
Unit
A
For the setting that is applied when *RST is sent, see Table A-16.
Response
Returns the OCP value in the NR3 form in response to CURR:PROT?.
•
When the phase input mode is changed to the single-phase input mode from the
three-phase input mode, the setting value of the output current and the
overcurrent protection (OCP) is limited to 75 % of the rated value of output
current and the overcurrent protection (OCP) of the three-phase input mode.
Summary of Contents for PAT160-25T
Page 17: ...General Description This chapter describes an overview and describes the features...
Page 22: ...1 6 PAT T 4kW...
Page 34: ...2 12 PAT T 4kW...
Page 106: ...6 12 PAT T 4kW...
Page 158: ...7 52 PAT T 4KW...
Page 159: ...Maintenance This chapter describes maintenance such as cleaning inspecting and calibrating...
Page 171: ...Specifications This chapter describes the specifications and gives the dimensions of the PAT...