Table 7-
8
Responses after sending VOLT 20;:VOLT:TRIG 10
VOLT:TRIG
Sets the voltage that is applied when INIT:SEQ1 / INIT:NAME TRAN and a
software trigger are sent.
Command
[SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPLitude]
{<numeric>|MIN|MAX}
[SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPLitude]?
[MIN|MAX]
Parameter
Val
u
e:
0 % to 105 % of the rated output voltage
(The default value is 0 % of the rated output voltage.)
Unit
V
For the setting that is applied when *RST is sent, see Table A-16.
Response
Returns the voltage value that is applied when a trigger is received in the NR3 form
in response to VOLT:TRIG?.
CURR:TRIG
Sets the current value that is applied when INIT:SEQ1 / INIT:NAME TRAN and a
software trigger are sent.
Command
[SOURce:]CURRent[:LEVel]:TRIGgered[:AMPLitude]
{<numeric>|MIN|MAX}
[SOURce:]CURRent[:LEVel]:TRIGgered[:AMPLitude]?
[MIN|MAX]
Parameter
Val
u
e:
0 % to 105 % of the rated output current
(The default value is 105 % of the rated output current.)
Unit
A
For the setting that is applied when *RST is sent, see Table A-16.
Response
Returns the current value that is applied when a trigger is received in the NR3 form
in response to CURR:TRIG?.
Response
VOLT?
VOLT:TRIG?
Immediately after the setting
20
10
After a trigger is sent
10
10
After *RST is sent
0
0
ABOR is sent before
sending a trigger
20
20 (cancel)
Voltage change VOLT 30 is
sent before sending a
trigger
30
30 (cancel)
Summary of Contents for PAT160-25T
Page 17: ...General Description This chapter describes an overview and describes the features...
Page 22: ...1 6 PAT T 4kW...
Page 34: ...2 12 PAT T 4kW...
Page 106: ...6 12 PAT T 4kW...
Page 158: ...7 52 PAT T 4KW...
Page 159: ...Maintenance This chapter describes maintenance such as cleaning inspecting and calibrating...
Page 171: ...Specifications This chapter describes the specifications and gives the dimensions of the PAT...