A-24
PAT-T 4kW
A.6.4
Status Monitoring
The PAT has two mandatory SCPI standard registers, STATus:OPERation and
STATus:QUEStionable, in addition to the IEEE488.2 standard registers.
Register basics
All SCPI registers have standard event/filter architecture, employing CONDition,
EVENt, ENABle, and optionally PTRansition and NTRansition. The CONDition
and EVENt are read-only registers working as status indicators, and the ENABle,
PTRansition and NTRansition are read-write registers working as event and
summary filters.
STATus:OPERation
The STATus:OPERation register records events or signals that occur during normal
operation.
For example, to check if the PAT is being regulated in CV state, check the CV bit
(bit 8) on the STATus:OPERation register.
:STATus:OPERation?
'
Check
w
hether the CV is set
STATus:QUEStionable
The STATus:QUEStionable register records events or signals that indicate abnormal
operation.
To check if the protection function is working, check the 0V bit (bit 0) on the
STATus: QUEStionable register.
:STATus:QUEStionable?
'
Check
w
hether the OV bit is set
See
See
See
Summary of Contents for PAT160-25T
Page 17: ...General Description This chapter describes an overview and describes the features...
Page 22: ...1 6 PAT T 4kW...
Page 34: ...2 12 PAT T 4kW...
Page 106: ...6 12 PAT T 4kW...
Page 158: ...7 52 PAT T 4KW...
Page 159: ...Maintenance This chapter describes maintenance such as cleaning inspecting and calibrating...
Page 171: ...Specifications This chapter describes the specifications and gives the dimensions of the PAT...