Dynamic current correction compensates for the current flowing into the output capacitor. The power
analyzer calculates what the additional current is and subtracts it from the current measurement,
thus providing accurate output current measurements. This feature is turned on by default and only
applies to the high current range.
Dynamic current correction increases the peak to peak noise in current measurements
on some power modules. It may also limit the measurement bandwidth as explained in
the next section. If either condition is a significant factor in your application, you should
turn dynamic current correction off.
To turn dynamic current correction on or off, press Scope View or Data Logger , then press Properties .
In the current range dropdown box, select the range labeled “CComp On” to turn current correction
on. Un-select the “CComp On” range to turn current correction off.
Keysight N6705C Operating and Service Guide
213
6 Advanced Source, Measurement, and Control
Summary of Contents for N6705C
Page 76: ......
Page 353: ...Status Diagram Keysight N6705C Operating and Service Guide 353 7 SCPI Programming Reference ...
Page 376: ...8 Verification and Calibration 376 Keysight N6705C Operating and Service Guide ...
Page 389: ...Keysight N6705C Operating and Service Guide 389 8 Verification and Calibration ...
Page 434: ......
Page 438: ...9 Service and Maintenance 438 Keysight N6705C Operating and Service Guide ...
Page 460: ...9 Service and Maintenance 460 Keysight N6705C Operating and Service Guide ...
Page 461: ...Keysight N6705C Operating and Service Guide 461 9 Service and Maintenance ...
Page 462: ...9 Service and Maintenance 462 Keysight N6705C Operating and Service Guide ...
Page 463: ...Keysight N6705C Operating and Service Guide 463 9 Service and Maintenance ...
Page 470: ...Mezzanine Parts 9 Service and Maintenance 470 Keysight N6705C Operating and Service Guide ...
Page 471: ...Front Panel Parts Keysight N6705C Operating and Service Guide 471 9 Service and Maintenance ...
Page 476: ......