Voltage Commands
Voltage commands program the output voltage and voltage protection functions. The SOURce
keyword is optional in the following commands.
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude] <
value
>, (@<chanlist>)
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude]? [MIN|MAX,] (@<chanlist>)
[SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPLitude] <
value
>, (@<chanlist>)
[SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPLitude]? [MIN|MAX,] (@<chanlist>)
Sets the immediate voltage level and the triggered voltage level when the output is in constant
voltage mode. The triggered level is a stored value that is transferred to the output when an output
Step is triggered.Units are in volts. The maximum value depends on the voltage rating of the unit.
Parameter
Typical Return
0 - maximum|MIN|MAX, *RST 0
The maximum value is 102% of the voltage rating of
the power module
<voltage level>
Sets the output voltage level to 20 V: VOLT 20, (@1)
Sets the triggered voltage level to 25 V: VOLT:TRIG 25, (@1)
The query returns the programmed voltage level in the form +n.nn for each channel
specified. Multiple responses are separated by commas.
[SOURce:]VOLTage:BWIDth LOW|HIGH1|HIGH2|HIGH3, (@<chanlist>)
[SOURce:]VOLTage:BWIDth? (@<chanlist>)
Sets the voltage bandwidth. This lets you optimize output response time with capacitive loads. Refer
to
for additional information
These compensation modes only apply when the unit is operating in voltage priority mode.
LOW - is optimized for stability with a wide range of output capacitors (from 0 to 150 μF).
HIGH1 - provides maximum up-programming speed and fastest transient response when the output
capacitance is restricted to 0–1 μF.
HIGH2 - provides maximum up-programming speed and fastest transient response when the output
capacitance is restricted to 0–7 μF.
HIGH3 - provides maximum up-programming speed and fastest transient response when the output
capacitance is restricted to 7–150 μF.
Parameter
Typical Return
LOW|HIGH1|HIGH2|HIGH3, *RST LOW
HIGH1, HIGH2, HIGH3, or LOW
Sets the voltage bandwidth to HIGH3: VOLT:BWID HIGH3, (@1)
Keysight N6705C Operating and Service Guide
343
7 SCPI Programming Reference
Summary of Contents for N6705C
Page 76: ......
Page 353: ...Status Diagram Keysight N6705C Operating and Service Guide 353 7 SCPI Programming Reference ...
Page 376: ...8 Verification and Calibration 376 Keysight N6705C Operating and Service Guide ...
Page 389: ...Keysight N6705C Operating and Service Guide 389 8 Verification and Calibration ...
Page 434: ......
Page 438: ...9 Service and Maintenance 438 Keysight N6705C Operating and Service Guide ...
Page 460: ...9 Service and Maintenance 460 Keysight N6705C Operating and Service Guide ...
Page 461: ...Keysight N6705C Operating and Service Guide 461 9 Service and Maintenance ...
Page 462: ...9 Service and Maintenance 462 Keysight N6705C Operating and Service Guide ...
Page 463: ...Keysight N6705C Operating and Service Guide 463 9 Service and Maintenance ...
Page 470: ...Mezzanine Parts 9 Service and Maintenance 470 Keysight N6705C Operating and Service Guide ...
Page 471: ...Front Panel Parts Keysight N6705C Operating and Service Guide 471 9 Service and Maintenance ...
Page 476: ......