INITiate:CONTinuous:TRANsient 0|OFF|1|ON, (@<chanlist>)
INITiate:CONTinuous:TRANsient? (@<chanlist>)
Continuously initiates the transient trigger system. This allows multiple triggers to generate multiple
output transients.
Parameter
Typical Return
0|OFF|1|ON , *RST ON
0 or 1
Continually initiates the trigger system: INIT:CONT:TRAN ON, (@1)
l
With continuous triggering disabled, the trigger system must be initiated for each trigger using the
INITiate:TRANsient command.
l
does not abort continuous triggers if INITiate:CONTinuous:TRANsient ON has
been programmed. In this case, the trigger system will automatically re-initiate.
Keysight N6705C Operating and Service Guide
295
7 SCPI Programming Reference
Summary of Contents for N6705C
Page 76: ......
Page 353: ...Status Diagram Keysight N6705C Operating and Service Guide 353 7 SCPI Programming Reference ...
Page 376: ...8 Verification and Calibration 376 Keysight N6705C Operating and Service Guide ...
Page 389: ...Keysight N6705C Operating and Service Guide 389 8 Verification and Calibration ...
Page 434: ......
Page 438: ...9 Service and Maintenance 438 Keysight N6705C Operating and Service Guide ...
Page 460: ...9 Service and Maintenance 460 Keysight N6705C Operating and Service Guide ...
Page 461: ...Keysight N6705C Operating and Service Guide 461 9 Service and Maintenance ...
Page 462: ...9 Service and Maintenance 462 Keysight N6705C Operating and Service Guide ...
Page 463: ...Keysight N6705C Operating and Service Guide 463 9 Service and Maintenance ...
Page 470: ...Mezzanine Parts 9 Service and Maintenance 470 Keysight N6705C Operating and Service Guide ...
Page 471: ...Front Panel Parts Keysight N6705C Operating and Service Guide 471 9 Service and Maintenance ...
Page 476: ......