Voltage Common Mode Rejection Ratio Calibration
Step 1. For this step, make the connections as shown in
. Connect an external jumper
between the +sense and the −sense terminals. Connect a second jumper from the +output to the
−sense terminal.
For models N6751A, N6752A, N6761A, and N6762A only, connect a 25
Ω
load resistor across the
+output and –output terminals.
This procedure is automatic and only takes a few seconds.
Front Panel Menu Reference
SCPI Command
From the Calibration menu, select Miscellaneous, Common Mode Rejection
Ratio, then Start. Then select Close.
After calibration completes, disconnect all jumpers and resistors.
Current Calibration
High Range Programming and Measurement
The output voltage may go negative at some point during this procedure on units that
have polarity reversal relays (Option 760).
Step 1. Connect a precision shunt resistor to the output. The shunt resistor should be able to measure
at least 120% of the output’s full-scale current (see
). Connect the Keysight 3458A DMM
across the shunt resistor.
Step 2. Select the full-scale current programming range. The value to program a range must be the
maximum current of the range. The following example selects the 5 A full-scale range of output 1.
Full-scale ranges vary according to model.
Front Panel Menu Reference
SCPI Command
From the Calibration menu, select Current, High Range Prog and Meas. Then
select Start.
Step 3. Select the first current calibration point and enter the data.
Front Panel Menu Reference
SCPI Command
Display shows: “Enter P1 measured data” Calculate the shunt current (I=V/R)
and enter the data.
Press Select when done.
CAL:LEV P1
*OPC?
CAL:DATA <data>
Step 4. Select the second current calibration point and enter the data.
8 Verification and Calibration
394
Keysight N6705C Operating and Service Guide
Summary of Contents for N6705C
Page 76: ......
Page 353: ...Status Diagram Keysight N6705C Operating and Service Guide 353 7 SCPI Programming Reference ...
Page 376: ...8 Verification and Calibration 376 Keysight N6705C Operating and Service Guide ...
Page 389: ...Keysight N6705C Operating and Service Guide 389 8 Verification and Calibration ...
Page 434: ......
Page 438: ...9 Service and Maintenance 438 Keysight N6705C Operating and Service Guide ...
Page 460: ...9 Service and Maintenance 460 Keysight N6705C Operating and Service Guide ...
Page 461: ...Keysight N6705C Operating and Service Guide 461 9 Service and Maintenance ...
Page 462: ...9 Service and Maintenance 462 Keysight N6705C Operating and Service Guide ...
Page 463: ...Keysight N6705C Operating and Service Guide 463 9 Service and Maintenance ...
Page 470: ...Mezzanine Parts 9 Service and Maintenance 470 Keysight N6705C Operating and Service Guide ...
Page 471: ...Front Panel Parts Keysight N6705C Operating and Service Guide 471 9 Service and Maintenance ...
Page 476: ......