[SOURce:]ARB:CURRent:STAircase:END:TIMe <
value
>, (@<chanlist>)
[SOURce:]ARB:CURRent:STAircase:END:TIMe? [MIN|MAX,] (@<chanlist>)
[SOURce:]ARB:VOLTage:STAircase:END:TIMe<
value
>, (@<chanlist>)
[SOURce:]ARB:VOLTage:STAircase:END:TIMe? [MIN|MAX,] (@<chanlist>)
Specifies the time in seconds, after the staircase completes, that the ending current or voltage level
persists. Referenced to t2 in the Staircase diagrams.
[SOURce:]ARB:SEQuence:STEP:CURRent:STAircase:END:TIMe <
value
>, <
step#
>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:CURRent:STAircase:END:TIMe? [MIN|MAX,] <
step#
>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:VOLTage:STAircase:END:TIMe <
value
>, <
step#
>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:VOLTage:STAircase:END:TIMe? [MIN|MAX,] <
step#
>, (@<chanlist>)
Specifies the time in seconds, after the staircase completes, that the ending current or voltage level
persists for Arb sequence steps.
Parameter
Typical Return
0 – 262.144 | MIN | MAX, *RST 0
<end time>
Step 0 - 99, *RST 0
(none)
Programs an end time after the current staircase:ARB:CURR:STA:END:TIM 2,(@1)
Programs an end time after the voltage staircase:ARB:VOLT:STA:END:TIM 2,(@1)
Programs an end time for step 0:ARB:SEQ:STEP:CURR:STA:END:TIM 2, 0,(@1)
[SOURce:]ARB:CURRent:STAircase:NSTeps <
value
>, (@<chanlist>)
[SOURce:]ARB:CURRent:STAircase:NSTeps? [MIN|MAX,] (@<chanlist>)
[SOURce:]ARB:VOLTage:STAircase:NSTeps<
value
>, (@<chanlist>)
[SOURce:]ARB:VOLTage:STAircase:NSTeps? [MIN|MAX,] (@<chanlist>)
Specifies the number of steps in the staircase.
[SOURce:]ARB:SEQuence:STEP:CURRent:STAircase:NSTeps <
value
>, <
step#
>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:CURRent:STAircase:NSTeps? [MIN|MAX,] <
step#
>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:VOLTage:STAircase:NSTeps <
value
>, <
step#
>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:VOLTage:STAircase:NSTeps? [MIN|MAX,] <
step#
>, (@<chanlist>)
Specifies the number of steps in the staircase for Arb sequence steps.
Parameter
Typical Return
0 – 500 | MIN | MAX, *RST 10
<number of steps>
Step 0 - 99, *RST 0
(none)
Programs five steps for the current staircase: ARB:CURR:STA:NST 5,(@1)
Programs five steps for the voltage staircase: ARB:VOLT:STA:NST 5,(@1)
Programs five steps for sequence step 0: ARB:SEQ:STEP::CURR:STA:NST 5, 0(@1)
7 SCPI Programming Reference
250
Keysight N6705C Operating and Service Guide
Summary of Contents for N6705C
Page 76: ......
Page 353: ...Status Diagram Keysight N6705C Operating and Service Guide 353 7 SCPI Programming Reference ...
Page 376: ...8 Verification and Calibration 376 Keysight N6705C Operating and Service Guide ...
Page 389: ...Keysight N6705C Operating and Service Guide 389 8 Verification and Calibration ...
Page 434: ......
Page 438: ...9 Service and Maintenance 438 Keysight N6705C Operating and Service Guide ...
Page 460: ...9 Service and Maintenance 460 Keysight N6705C Operating and Service Guide ...
Page 461: ...Keysight N6705C Operating and Service Guide 461 9 Service and Maintenance ...
Page 462: ...9 Service and Maintenance 462 Keysight N6705C Operating and Service Guide ...
Page 463: ...Keysight N6705C Operating and Service Guide 463 9 Service and Maintenance ...
Page 470: ...Mezzanine Parts 9 Service and Maintenance 470 Keysight N6705C Operating and Service Guide ...
Page 471: ...Front Panel Parts Keysight N6705C Operating and Service Guide 471 9 Service and Maintenance ...
Page 476: ......