Peak Current Limit Calibration
For models N6753A through N6756A and N6763A through N6766A, the peak current
limit calibration is valid for five years.
Step 1. Connect only the electronic load or equivalent resistor (up to 500 W rating) across the output
terminals for the first calibration point.
Model
1st calibration point
2nd calibration point
N6751A/N6752A
N6761A/N6762A
25
Ω
resistor (do not use electronic load)
4
Ω
resistor (do not use electronic load)
N6753A/N6763A
Electronic load set to 26.25A, CC mode (0.6
Ω
equi-
valent, including lead resistance)
Electronic load set to 13.12A, CC mode (1.2
Ω
equi-
valent, including lead resistance)
N6754A/N6764A
Electronic load set to 10.5A, CC mode (3.8
Ω
equi-
valent, including lead resistance)
Electronic load set to 5.25A, CC mode (7.6
Ω
equi-
valent, including lead resistance)
N6755A/N6765A
Electronic load set to 43.75A, CC mode (0.37
Ω
equi-
valent, including lead resistance)
Electronic load set to 21.88A, CC mode (0.73
Ω
equi-
valent, including lead resistance)
N6756A/N6766A
Electronic load set to 17.5A, CC mode (2.3
Ω
equi-
valent, including lead resistance)
Electronic load set to 8.75A, CC mode (8.75
Ω
equi-
valent, including lead resistance)
Step 2. Select peak current limit calibration.
Front Panel Menu Reference
SCPI Command
From the Calibration menu, select Miscellaneous, Peak Current Limit. Then
select Start.
Step 3. Select the first current calibration point.
Front Panel Menu Reference
SCPI Command
Select Next. The calibration is automatic.
Step 4. Set the electronic load or connect the appropriate resistor for the second calibration point.
Step 5. Select the second current calibration point.
Front Panel Menu Reference
SCPI Command
Select Next. The calibration is automatic.
Downprogrammer Calibration
Step 1. Remove all loads from the output. This procedure is automatic and takes a few seconds.
Keysight N6705C Operating and Service Guide
397
8 Verification and Calibration
Summary of Contents for N6705C
Page 76: ......
Page 353: ...Status Diagram Keysight N6705C Operating and Service Guide 353 7 SCPI Programming Reference ...
Page 376: ...8 Verification and Calibration 376 Keysight N6705C Operating and Service Guide ...
Page 389: ...Keysight N6705C Operating and Service Guide 389 8 Verification and Calibration ...
Page 434: ......
Page 438: ...9 Service and Maintenance 438 Keysight N6705C Operating and Service Guide ...
Page 460: ...9 Service and Maintenance 460 Keysight N6705C Operating and Service Guide ...
Page 461: ...Keysight N6705C Operating and Service Guide 461 9 Service and Maintenance ...
Page 462: ...9 Service and Maintenance 462 Keysight N6705C Operating and Service Guide ...
Page 463: ...Keysight N6705C Operating and Service Guide 463 9 Service and Maintenance ...
Page 470: ...Mezzanine Parts 9 Service and Maintenance 470 Keysight N6705C Operating and Service Guide ...
Page 471: ...Front Panel Parts Keysight N6705C Operating and Service Guide 471 9 Service and Maintenance ...
Page 476: ......