To select Digital pin triggers, where n is the pin number that will generate the trigger:
To select a Transient output as trigger, where n is the output channel that will generate the trigger
signal:
To enable trigger signals for output Steps or Lists: (Specify if the list trigger occurs at the Beginning Of
the Step, or End Of the Step.)
STEP:TOUT ON,(@3)
LIST:TOUT:BOST 1, (@3)
LIST:TOUT:EOST 1, (@3)
To select a voltage or current level of another output as the trigger for output 1: (Output 3 will
generate the voltage or current level to trigger output 1.)
TRIG:ACQ:SOUR VOLT3,(@1)
TRIG:ACQ:SOUR CURR3,(@1)
To specify the voltage or current trigger level and slope of output 3:
TRIG:ACQ:CURR 10,(@3)
TRIG:ACQ:CURR:SLOP POS,(@3)
TRIG:ACQ:VOLT 10,(@3)
TRIG:ACQ:VOLT:SLOP POS,(@3)
A non-programmable hysteresis for level triggers reduces false triggering on slow
signals. The hysteresis is set to 0.0005 X maximum range value. For example, on the 50
V range, the hysteresis is approximately 25 mV.
Initiate the Measurement
When the power analyzer is turned on, the trigger system is in the idle state. In this state, the trigger
system is disabled, ignoring all triggers. The INITiate command enables the measurement system to
receive triggers. To initiate the trigger system for all four outputs:
It takes a few milliseconds for the instrument to be ready to receive a trigger signal after receiving the
INITiate:ACQuire command, and it can take longer for Keysight Models N678xA SMU.
If a trigger occurs before the trigger system is ready for it, the trigger will be ignored. You can test the
WTG_meas bit in the operation status register to know when the instrument is ready to receive a
trigger after being initiated.
If a bit value of 8 is returned in the query, the WTG_meas bit is true, and the instrument is ready to
receive the trigger signal. Refer to
for more information.
6 Advanced Source, Measurement, and Control
204
Keysight N6705C Operating and Service Guide
Summary of Contents for N6705C
Page 76: ......
Page 353: ...Status Diagram Keysight N6705C Operating and Service Guide 353 7 SCPI Programming Reference ...
Page 376: ...8 Verification and Calibration 376 Keysight N6705C Operating and Service Guide ...
Page 389: ...Keysight N6705C Operating and Service Guide 389 8 Verification and Calibration ...
Page 434: ......
Page 438: ...9 Service and Maintenance 438 Keysight N6705C Operating and Service Guide ...
Page 460: ...9 Service and Maintenance 460 Keysight N6705C Operating and Service Guide ...
Page 461: ...Keysight N6705C Operating and Service Guide 461 9 Service and Maintenance ...
Page 462: ...9 Service and Maintenance 462 Keysight N6705C Operating and Service Guide ...
Page 463: ...Keysight N6705C Operating and Service Guide 463 9 Service and Maintenance ...
Page 470: ...Mezzanine Parts 9 Service and Maintenance 470 Keysight N6705C Operating and Service Guide ...
Page 471: ...Front Panel Parts Keysight N6705C Operating and Service Guide 471 9 Service and Maintenance ...
Page 476: ......