*SAV <0-9>
Save State
Saves the present state of the instrument to one of ten non-volatile memory locations.
Any state previously stored in the same location will be overwritten. Refer to
for a list of
instrument settings that can be saved. Settings
Parameter
Typical Return
0 - 9
(none)
Save state to location 1: *SAV 1
l
If a particular state is desired at power-on, it should be stored in location 0. Location 0 is auto-
matically recalled at power turn-on when the Output Power-On state is set to RCL 0.
l
The calibration state is NOT saved as part of the *SAV operation.
l
Data saved in
is not affected by the *SAV command.
l
A reset (*RST) command does not affect the instrument states stored in memory. Once a state is
stored, it remains until it is overwritten or specifically deleted.
*SRE <
value
>
*SRE?
Service Request Enable Command and Query
. Sets the value of the Service Request Enable register.
This determines which bits from the
are summed to set the Master Status
Summary (MSS) bit and the Request for Service (RQS) summary bit. A 1 in any Service Request Enable
register bit position enables the corresponding Status Byte register bit. All such enabled bits are then
logically OR-ed to cause the MSS bit of the Status Byte register to be set. Refer to
for
more information.
The query reads the enable register and returns a decimal value which corresponds to the binary-
weighted sum of all bits set in the register.
Parameter
Typical Return
A decimal value that corresponds to the binary-weighted sum of the bits
in the register.
<bit value>
Enable bit 3 and 4 in the enable register: *SRE 24
l
When a serial poll is conducted in response to a service request, RQS is cleared but MSS is not.
l
When *SRE is cleared (by programming it with 0), the power analyzer cannot generate a service
request.
*STB?
Status Byte Query
. Reads the
, which contains the status summary bits and the
Output Queue MAV bit. The Status Byte is a read-only register and the bits are NOT cleared when it is
read. Refer to
for more information.
7 SCPI Programming Reference
292
Keysight N6705C Operating and Service Guide
Summary of Contents for N6705C
Page 76: ......
Page 353: ...Status Diagram Keysight N6705C Operating and Service Guide 353 7 SCPI Programming Reference ...
Page 376: ...8 Verification and Calibration 376 Keysight N6705C Operating and Service Guide ...
Page 389: ...Keysight N6705C Operating and Service Guide 389 8 Verification and Calibration ...
Page 434: ......
Page 438: ...9 Service and Maintenance 438 Keysight N6705C Operating and Service Guide ...
Page 460: ...9 Service and Maintenance 460 Keysight N6705C Operating and Service Guide ...
Page 461: ...Keysight N6705C Operating and Service Guide 461 9 Service and Maintenance ...
Page 462: ...9 Service and Maintenance 462 Keysight N6705C Operating and Service Guide ...
Page 463: ...Keysight N6705C Operating and Service Guide 463 9 Service and Maintenance ...
Page 470: ...Mezzanine Parts 9 Service and Maintenance 470 Keysight N6705C Operating and Service Guide ...
Page 471: ...Front Panel Parts Keysight N6705C Operating and Service Guide 471 9 Service and Maintenance ...
Page 476: ......