A percent of 0 means the trigger occurs at the beginning of the running datalog, while 100 means that
the trigger occurs at the end of the running datalog. Any value between 0 and 100 can be set.
Parameter
Typical Return
0 - 100, *RST 0
<offset percent>
Specifies a trigger offset of 50%: SENS:DLOG:OFFS 50
l
The offset must be set to zero when interleaved data logging is in effect.
SENSe:DLOG:PERiod <
value
>
SENSe:DLOG:PERiod? [MIN|MAX,]
Sets the integration time of the datalog measurement.
Although the absolute minimum logging period is 102.4 microseconds, the actual minimum varies as a
function of the number of readings that are being logged (see
).
This command replaces the previous SENSe:DLOG:TINTerval command and should be
used in new applications. SENSe:DLOG:TINTerval is still available for backward
compatibility.
Parameter
Typical Return
0.0002048 to 60|MIN|MAX, *RST 0.1
<datalog period>
Specifies a datalog period of 0.01 seconds: SENS:DLOG:PER 0.01, (@1)
l
This function cannot be set to values less than 75 milliseconds when interleaved data logging is in
effect.
SENSe:DLOG:TIME <
value
>
SENSe:DLOG:TIME? [MIN|MAX,]
Specifies the duration of the datalog in seconds. Values from 1 second up to the equivalent of 99,999
hours can be specified.
The resulting datalog file cannot exceed 2E9 bytes (1.87 Gbytes in Microsoft Windows units). If the
duration value causes the file to exceeds this size, the time interval will automatically be adjusted to a
longer interval to reduce the file size
Parameter
Typical Return
1 to 359,996,400|MIN|MAX, *RST 30
<datalog time>
Specifies a datalog duration of 5000 seconds: SENS:DLOG:TIME 5000
Keysight N6705C Operating and Service Guide
319
7 SCPI Programming Reference
Summary of Contents for N6705C
Page 76: ......
Page 353: ...Status Diagram Keysight N6705C Operating and Service Guide 353 7 SCPI Programming Reference ...
Page 376: ...8 Verification and Calibration 376 Keysight N6705C Operating and Service Guide ...
Page 389: ...Keysight N6705C Operating and Service Guide 389 8 Verification and Calibration ...
Page 434: ......
Page 438: ...9 Service and Maintenance 438 Keysight N6705C Operating and Service Guide ...
Page 460: ...9 Service and Maintenance 460 Keysight N6705C Operating and Service Guide ...
Page 461: ...Keysight N6705C Operating and Service Guide 461 9 Service and Maintenance ...
Page 462: ...9 Service and Maintenance 462 Keysight N6705C Operating and Service Guide ...
Page 463: ...Keysight N6705C Operating and Service Guide 463 9 Service and Maintenance ...
Page 470: ...Mezzanine Parts 9 Service and Maintenance 470 Keysight N6705C Operating and Service Guide ...
Page 471: ...Front Panel Parts Keysight N6705C Operating and Service Guide 471 9 Service and Maintenance ...
Page 476: ......