N6781A, '82A, and '84A
Report Number ______
Date__________
Test Description
Model
Min. Specs
Results
Max. Specs
CC Load Effect
3 A output range:
1 A output range:
300 mA output range:
100 mA & 10 mA output ranges:
All
All
N6781A/N6782A
N6784A
– 100 µA
– 50 µA
– 50 µA
– 1 µA
_______
_______
_______
_______
+ 100 µA
+ 50 µA
+ 50 µA
+ 1 µA
N6781A Resistance Tests
Resistance Programming
20 V output range:
6 V output range:
N6781A
N6781A
0.996 Ω
0.9975 Ω
_______
_______
1.004 Ω
1.0025 Ω
Test Description
N6781A, '82A, and '84A Settings
Voltage Programming & Readback Min Voltage:
Voltage Programming & Readback High Voltage:
Voltage Programming & Readback Low Voltage:
600 mV Programming & 100 mV Readback
CV Load Effect, 20 V range
CV Load Effect, 6 V range
CV Load Effect, 600 mV range
CV Ripple and Noise, 20 V range
CV Ripple and Noise, 6 V range
Transient Response, 6 V range
Transient Response, 20 V range
Current Programming & Readback, Min Current
Current Programming & Readback, High Current
Current Programming 300 mA & 100 mA ranges
Current Programming 10 mA range
Current Readback 100 mA range
Current Readback 1 mA Range
Current Readback 10 μA Range
CC Load Effect, 3 A range
CC Load Effect, 1 A range
CC Load Effect, 300 mA range
CC Load Effect, 100 mA range
CC Load Effect, 10 mA range
N6781A Resistance Programming, 20 V range
N6781A Resistance Programming, 6 V range
Voltage Priority; 0 V, 1 A
Voltage Priority; 20 V, 1 A
Voltage Priority; 1 V, 1 A
Voltage Priority; 0.1 V, 1 A
Voltage Priority; 20 V, 1 A
Voltage Priority; 6 V, 3 A
Voltage Priority; 100 mV, 3 A
Voltage Priority; 20 V, 1 A
Voltage Priority; 6 V, 3 A
Voltage Priority; Comp.= HIGH3; 6V, from 0.1 to 1.5 A
Voltage Priority; Comp.= HIGH3; 20V, from 0.1 to 0.9 A
Current Priority; 0 A, 6 V
Current Priority; 3 A, 6 V
Current Priority; 100 mA, 20 V
Current Priority; 10 mA, 20 V
Voltage Priority; 100 mA, 1 V
Voltage Priority; 1 mA, 10 V
Voltage Priority; 10 μA, 10 V
Current Priority; 3 A, 6 V
Current Priority; 1 A, 6 V
Current Priority; 100 mA, 6 V
Current Priority; 100 mA, 6 V
Current Priority; 10 mA, 6 V
Voltage Priority; 10 V, 1 Ω (Elect. load: CC mode, 20 V, 1A)
Voltage Priority; 6 V, 1 Ω (Elect. load: CC mode, 20 V, 3A)
Keysight N6705C Operating and Service Guide
431
8 Verification and Calibration
Summary of Contents for N6705C
Page 76: ......
Page 353: ...Status Diagram Keysight N6705C Operating and Service Guide 353 7 SCPI Programming Reference ...
Page 376: ...8 Verification and Calibration 376 Keysight N6705C Operating and Service Guide ...
Page 389: ...Keysight N6705C Operating and Service Guide 389 8 Verification and Calibration ...
Page 434: ......
Page 438: ...9 Service and Maintenance 438 Keysight N6705C Operating and Service Guide ...
Page 460: ...9 Service and Maintenance 460 Keysight N6705C Operating and Service Guide ...
Page 461: ...Keysight N6705C Operating and Service Guide 461 9 Service and Maintenance ...
Page 462: ...9 Service and Maintenance 462 Keysight N6705C Operating and Service Guide ...
Page 463: ...Keysight N6705C Operating and Service Guide 463 9 Service and Maintenance ...
Page 470: ...Mezzanine Parts 9 Service and Maintenance 470 Keysight N6705C Operating and Service Guide ...
Page 471: ...Front Panel Parts Keysight N6705C Operating and Service Guide 471 9 Service and Maintenance ...
Page 476: ......