Select the Histogram Trigger Source
The TRIGger:HISTogram command generates an immediate trigger regardless of the trigger source.
Unless you are using this command, select a trigger source from the following:
BUS
Selects GPIB device trigger, *TRG, or <GET> (Group Execute Trigger).
EXTernal
Selects the trigger input BNC connector. Requires a low-true trigger signal.
IMMediate
Selects the immediate trigger source. This triggers the histogram immediately when initiated.
Pin<1-7>
Selects a specific pin that is configured as a Trigger Input on the digital port. The selected pin must be
configured as a Trigger Input in order to be used as a trigger source (see
).
Use the following commands to select a trigger source. To select Bus triggers for output 1:
To select the IMMediate trigger source for output 1:
To select the EXTernal trigger source for output 1:
To select Digital pin triggers on pin 3 for output 1:
Initiate and Trigger the Histogram
When the power analyzer is turned on, the trigger system is in the Idle state. In this state, the trigger
system is disabled, ignoring all triggers. The INITiate command enables the measurement system to
receive triggers. To initiate and then trigger histogram measurements on output 1:
If the trigger source is BUS, you can also program *TRG or <GET> to trigger the histogram.
When a histogram measurement is initiated and triggered, the current is continuously sampled.
Histograms sample at the rate of 20.48 microseconds per sample. Each sampled value is compared to
the amplitude range of each bin of the histogram and the value of the bin that the sampled value falls
within is incremented by 1. There is some overlap between the lowest bins of the high range and the
highest bins of the low range; values that overlap both ranges are placed in the low range buckets.
Each Fetch returns the latest cumulative histogram data. The measurement continues until it is
aborted. The count bins are 64-bits wide, which eliminates any concern for overflow.
Terminate the Measurement
Measurements continue until the histogram is aborted. To abort the histogram measurement:
6 Advanced Source, Measurement, and Control
210
Keysight N6705C Operating and Service Guide
Summary of Contents for N6705C
Page 76: ......
Page 353: ...Status Diagram Keysight N6705C Operating and Service Guide 353 7 SCPI Programming Reference ...
Page 376: ...8 Verification and Calibration 376 Keysight N6705C Operating and Service Guide ...
Page 389: ...Keysight N6705C Operating and Service Guide 389 8 Verification and Calibration ...
Page 434: ......
Page 438: ...9 Service and Maintenance 438 Keysight N6705C Operating and Service Guide ...
Page 460: ...9 Service and Maintenance 460 Keysight N6705C Operating and Service Guide ...
Page 461: ...Keysight N6705C Operating and Service Guide 461 9 Service and Maintenance ...
Page 462: ...9 Service and Maintenance 462 Keysight N6705C Operating and Service Guide ...
Page 463: ...Keysight N6705C Operating and Service Guide 463 9 Service and Maintenance ...
Page 470: ...Mezzanine Parts 9 Service and Maintenance 470 Keysight N6705C Operating and Service Guide ...
Page 471: ...Front Panel Parts Keysight N6705C Operating and Service Guide 471 9 Service and Maintenance ...
Page 476: ......