Front Panel Menu Reference
SCPI Command
From the Calibration menu, select Current, Current Limit. Then select Start.
For N6781A, N6782A, N6784A:
For N6785A, N6786A:
Step 3. Select the first current calibration point and enter the data.
Front Panel Menu Reference
SCPI Command
Display shows: “Enter P1 measured data” Calculate the shunt current (I=V/R)
and enter the data. Press Select when done.
CAL:LEV P1
*OPC?
CAL:DATA <data>
Step 4. Select the second current calibration point and enter the data. Wait a minimum of 1 minute for
the output current to stabilize before entering the data.
Front Panel Menu Reference
SCPI Command
Wait1 minute and select Next.
Display shows: “Enter P2 measured data”
Calculate the shunt current (I=V/R) and enter the data. Press Select when done.
Press Next to finish calibration.
CAL:LEV P2
*OPC?
CAL:DATA <data>
Resistance Calibration - Models N6781A, N6785A
Step 1. Connect an electronic load directly to the output terminals. Also connect the voltage input of
the Keysight 3458A to the output terminals (see
). Set the load to constant current mode,
current to 0.95A.
Step 2. Select the 20 V resistance range of output 1.
Front Panel Menu Reference
SCPI Command
From the Calibration menu, select Miscellaneous Resistance, 20 V Range. Then
select Start.
Step 3. Select the first resistance calibration point and enter the data.
Front Panel Menu Reference
SCPI Command
Display shows: “Enter P1 measured data” Enter the data from the external DMM.
Press Select when done.
CAL:LEV P1
*OPC?
CAL:DATA <data>
Step 4. Select the second resistance calibration point and enter the data.
Front Panel Menu Reference
SCPI Command
Display shows: “Enter P2 measured data” Enter the data from the external DMM.
Press Select when done.
CAL:LEV P2
*OPC?
CAL:DATA <data>
Keysight N6705C Operating and Service Guide
407
8 Verification and Calibration
Summary of Contents for N6705C
Page 76: ......
Page 353: ...Status Diagram Keysight N6705C Operating and Service Guide 353 7 SCPI Programming Reference ...
Page 376: ...8 Verification and Calibration 376 Keysight N6705C Operating and Service Guide ...
Page 389: ...Keysight N6705C Operating and Service Guide 389 8 Verification and Calibration ...
Page 434: ......
Page 438: ...9 Service and Maintenance 438 Keysight N6705C Operating and Service Guide ...
Page 460: ...9 Service and Maintenance 460 Keysight N6705C Operating and Service Guide ...
Page 461: ...Keysight N6705C Operating and Service Guide 461 9 Service and Maintenance ...
Page 462: ...9 Service and Maintenance 462 Keysight N6705C Operating and Service Guide ...
Page 463: ...Keysight N6705C Operating and Service Guide 463 9 Service and Maintenance ...
Page 470: ...Mezzanine Parts 9 Service and Maintenance 470 Keysight N6705C Operating and Service Guide ...
Page 471: ...Front Panel Parts Keysight N6705C Operating and Service Guide 471 9 Service and Maintenance ...
Page 476: ......