Test Equipment and Setups
Equipment Required
The test equipment recommended for the verification and calibration procedures is listed below. If the
exact instrument is not available, substitute calibration standards of equivalent accuracy.
Instrument
Requirements
Recommended Model
Use1
Digital Multimeter
Resolution: 10 nV @ 1V
Readout: 8 1/2 digits
Accuracy: 20 ppm
Keysight 3458A or equivalent
V, C
Current Shunt2
15 A (0.1
Ω
)
30 A (0.1
Ω
)
50 A (0.05
Ω
)
Guildline 9230A-15R
Guildline 9230A-30
Guildline 9230A-50
V, C
Load Resistors
25
Ω
, 250 W, 1% (N6751A/52A/61A/62A)
4
Ω
, 250 W, 1% (N6751A/52A/61A/62A)
20 k
Ω
, 1%, 1/8W (for option 1UA/2UA)
10
Ω
, 1%, 1/8W (N678xA SMU)
10 k
Ω
, 1%, 1/8W (N678xA SMU)
1 M
Ω
, 1%, 1/8W (N678xA SMU)
Vishay NH-250-25
Ω
-1%
Vishay NH-250-4
Ω
-1%
C
C
V, C
V, C
V, C
V, C
Electronic load2
150 V, 50 A minimum, with transient capability and a slew
rate of 833 kA/s or better
Keysight N3300A mainframe, with
N330xA modules
V, C
Keysight N6784A
Load for N678xA SMU and N6783A (set to CC load mode)
Keysight N6784A
V
GPIB controller
Full GPIB capabilities
Keysight 82350B or equivalent
V, C
Oscilloscope
Sensitivity: 1 mV
Bandwidth Limit: 20 MHz
Probe: 1:1 with RF tip
Keysight DSO6054A or equivalent
V
RMS Voltmeter
True RMS
Bandwidth: 20 MHz
Sensitivity: 100 µV
Rhode and Schwartz Model URE3
or equivalent
V
Differential Amplifier
Bandwidth: 20 MHz
LeCroy DA1855A, DA1850A, or
equivalent
V
Terminations
1 – 50
Ω
BNC termination
2 – 50
Ω
1/8 W termination resistors
V
Variable voltage transformer
or AC source
Adjustable to highest rated input voltage
Power: 500 VA
Keysight 6813B or equivalent
V
RC – for N678xA transient
test
Capacitor: 150 μF fixed film
Resistor 50 m
Ω
EPCOS B32526R0157K
Vishay LVR01R0500FE70
V
1 V = Verification C = Calibration
2 Dependent on rating of model being tested
8 Verification and Calibration
372
Keysight N6705C Operating and Service Guide
Summary of Contents for N6705C
Page 76: ......
Page 353: ...Status Diagram Keysight N6705C Operating and Service Guide 353 7 SCPI Programming Reference ...
Page 376: ...8 Verification and Calibration 376 Keysight N6705C Operating and Service Guide ...
Page 389: ...Keysight N6705C Operating and Service Guide 389 8 Verification and Calibration ...
Page 434: ......
Page 438: ...9 Service and Maintenance 438 Keysight N6705C Operating and Service Guide ...
Page 460: ...9 Service and Maintenance 460 Keysight N6705C Operating and Service Guide ...
Page 461: ...Keysight N6705C Operating and Service Guide 461 9 Service and Maintenance ...
Page 462: ...9 Service and Maintenance 462 Keysight N6705C Operating and Service Guide ...
Page 463: ...Keysight N6705C Operating and Service Guide 463 9 Service and Maintenance ...
Page 470: ...Mezzanine Parts 9 Service and Maintenance 470 Keysight N6705C Operating and Service Guide ...
Page 471: ...Front Panel Parts Keysight N6705C Operating and Service Guide 471 9 Service and Maintenance ...
Page 476: ......