Low Range Voltage Programming - Models N6761A through N6766A
Step 1. Select the low-voltage programming range. The value entered must be the maximum voltage
of the range to be calibrated. This example selects the 5 V range of output 1.
Front Panel Menu Reference
SCPI Command
From the Calibration menu, select Voltage, Low Range Programming. Then select
Start.
Step 2. Select the first voltage calibration point and enter the data. This value may be positive or
negative.
Front Panel Menu Reference
SCPI Command
Display shows: “Enter P1 measured data”Enter the data from the external DMM.
Press Select when done.
CAL:LEV P1
*OPC?
CAL:DATA <data>
Step 3. Select the second voltage calibration point and enter the data.
Front Panel Menu Reference
SCPI Command
Display shows: “Enter P2 measured data”Enter the data from the external
DMM.Press Select when done.
Press Next to finish calibration.
CAL:LEV P2
*OPC?
CAL:DATA <data>
Low Range Voltage Measurement - Models N6761A through N6766A
Step 1. Select the low-voltage measurement range. The value entered must be the maximum voltage
of the range to be calibrated. This example selects the 5 V range of output 1.
Front Panel Menu Reference
SCPI Command
From the Calibration menu, select Voltage, Low Range Measurement. Then select
Start.
Step 2. Select the first voltage calibration point and enter the data.
Front Panel Menu Reference
SCPI Command
Display shows: “Enter P1 measured data”Enter the data from the external DMM.
Press Select when done.
CAL:LEV P1
*OPC?
CAL:DATA <data>
Step 3. Select the second voltage calibration point and enter the data.
Front Panel Menu Reference
SCPI Command
Display shows: “Enter P2 measured data”Enter the data from the external DMM.
Press Select when done.
Press Next to finish calibration.
CAL:LEV P2
*OPC?
CAL:DATA <data>
Keysight N6705C Operating and Service Guide
393
8 Verification and Calibration
Summary of Contents for N6705C
Page 76: ......
Page 353: ...Status Diagram Keysight N6705C Operating and Service Guide 353 7 SCPI Programming Reference ...
Page 376: ...8 Verification and Calibration 376 Keysight N6705C Operating and Service Guide ...
Page 389: ...Keysight N6705C Operating and Service Guide 389 8 Verification and Calibration ...
Page 434: ......
Page 438: ...9 Service and Maintenance 438 Keysight N6705C Operating and Service Guide ...
Page 460: ...9 Service and Maintenance 460 Keysight N6705C Operating and Service Guide ...
Page 461: ...Keysight N6705C Operating and Service Guide 461 9 Service and Maintenance ...
Page 462: ...9 Service and Maintenance 462 Keysight N6705C Operating and Service Guide ...
Page 463: ...Keysight N6705C Operating and Service Guide 463 9 Service and Maintenance ...
Page 470: ...Mezzanine Parts 9 Service and Maintenance 470 Keysight N6705C Operating and Service Guide ...
Page 471: ...Front Panel Parts Keysight N6705C Operating and Service Guide 471 9 Service and Maintenance ...
Page 476: ......