100 mA Measurement Range - All Models
Step 1. Connect only the current measurement terminals of the Keysight 3458A to the output
terminals (see
without the resistor).
Step 2. Select the 100 mA measurement range of output 1.
Front Panel Menu Reference
SCPI Command
From the Calibration menu, select Current, 100 mA Measurement Range. Then
select Start.
Step 3. Select the first current calibration point and enter the data.
Front Panel Menu Reference
SCPI Command
Display shows: “Enter P1 measured data” Measure the current with the Keysight
3458A and enter the data.
Press Select when done.
CAL:LEV P1
*OPC?
CAL:DATA <data>
Step 4. Select the second current calibration point for the negative range and enter the data. The
value should be negative.
Front Panel Menu Reference
SCPI Command
Display shows: “Enter P2 measured data”
Measure the current with the Keysight 3458A and enter the data. Press Select
when done.
CAL:LEV P2
*OPC?
CAL:DATA <data>
1 mA Measurement Range - All Models
Step 1. Connect the Keysight 3458A in series with a 10 k
Ω
resistor to the output (see
and
).
Step 2. Select the 1 mA measurement range of output 1.
Front Panel Menu Reference
SCPI Command
From the Calibration menu, select Current, 1 mA Measurement Range. Then
select Start.
Step 3. Select the first current calibration point and enter the data.
Front Panel Menu Reference
SCPI Command
Display shows: “Enter P1 measured data” Measure the current with the Keysight
3458A and enter the data.
Press Select when done.
CAL:LEV P1
*OPC?
CAL:DATA <data>
Step 4 applies only to models N6781A, N6782A, and N6784A
Step 4. Select the second current calibration point for the negative range and enter the data. The
value should be negative.
Keysight N6705C Operating and Service Guide
405
8 Verification and Calibration
Summary of Contents for N6705C
Page 76: ......
Page 353: ...Status Diagram Keysight N6705C Operating and Service Guide 353 7 SCPI Programming Reference ...
Page 376: ...8 Verification and Calibration 376 Keysight N6705C Operating and Service Guide ...
Page 389: ...Keysight N6705C Operating and Service Guide 389 8 Verification and Calibration ...
Page 434: ......
Page 438: ...9 Service and Maintenance 438 Keysight N6705C Operating and Service Guide ...
Page 460: ...9 Service and Maintenance 460 Keysight N6705C Operating and Service Guide ...
Page 461: ...Keysight N6705C Operating and Service Guide 461 9 Service and Maintenance ...
Page 462: ...9 Service and Maintenance 462 Keysight N6705C Operating and Service Guide ...
Page 463: ...Keysight N6705C Operating and Service Guide 463 9 Service and Maintenance ...
Page 470: ...Mezzanine Parts 9 Service and Maintenance 470 Keysight N6705C Operating and Service Guide ...
Page 471: ...Front Panel Parts Keysight N6705C Operating and Service Guide 471 9 Service and Maintenance ...
Page 476: ......