Step 1. Turn off the power analyzer and connect a load resistor or electronic load, differential
amplifier, and an oscilloscope (ac coupled) to the output (see
).
Step 2.
For models N6751A/52A, N6761A/62A, and N678xA SMU
- use an appropriate load resistor to
keep the power analyzer at the instrument setting specified in the test record form under “CV Ripple
and Noise”. For all other power modules - use an electronic load to keep the power analyzer at the
instrument setting specified in the test record form under “CV Ripple and Noise”.
Step 3. As shown in the diagram, use two BNC cables to connect the differential amplifier to the + and
− output terminals. Each cable should be terminated by a 50
Ω
resistor. The shields of the two BNC
cables should be connected together. Connect the output of the differential amplifier to the
oscilloscope with a 50
Ω
termination at the input of the oscilloscope.
Step 4. Set the differential amplifier to multiply by ten, divide by one, and 1 M
Ω
input resistance. The
positive and negative inputs of the differential amplifier should be set to AC coupling. Set the
oscilloscope’s time base to 5 ms/div, and the vertical scale to 10 mV/div. Turn the bandwidth limit on
(usually 20 or 30 MHz), and set the sampling mode to peak detect.
Step 5. Program the power analyzer to the settings indicated in the in the test record form for the
appropriate model under “CV Ripple and Noise” and enable the output. Let the oscilloscope run for a
few seconds to generate enough measurement points. On the Keysight Infiniium scope, the maximum
peak-to-peak voltage measurement is indicated at the bottom of the screen on the right side. Divide
this value by 10 to get the CV peak-to-peak noise measurement. The result should not exceed the
peak-to-peak limits in the test record form for the appropriate model under “CV Ripple and Noise,
peak-to-peak”.
If the measurement contains any question marks, clear the measurement and try again.
This means that some of the scope data received was questionable.
Step 6. Disconnect the oscilloscope and connect an rms voltmeter in its place. Do not disconnect the
50 ohm termination. Divide the reading of the rms voltmeter by 10. The result should not exceed the
rms limits in the test record form for the appropriate model under “CV Ripple and Noise, rms”.
Step 7.
For models N6781A, N6782A, and N6784A only.
Repeat steps 2 through 6 for the 6 V range.
Keysight N6705C Operating and Service Guide
379
8 Verification and Calibration
Summary of Contents for N6705C
Page 76: ......
Page 353: ...Status Diagram Keysight N6705C Operating and Service Guide 353 7 SCPI Programming Reference ...
Page 376: ...8 Verification and Calibration 376 Keysight N6705C Operating and Service Guide ...
Page 389: ...Keysight N6705C Operating and Service Guide 389 8 Verification and Calibration ...
Page 434: ......
Page 438: ...9 Service and Maintenance 438 Keysight N6705C Operating and Service Guide ...
Page 460: ...9 Service and Maintenance 460 Keysight N6705C Operating and Service Guide ...
Page 461: ...Keysight N6705C Operating and Service Guide 461 9 Service and Maintenance ...
Page 462: ...9 Service and Maintenance 462 Keysight N6705C Operating and Service Guide ...
Page 463: ...Keysight N6705C Operating and Service Guide 463 9 Service and Maintenance ...
Page 470: ...Mezzanine Parts 9 Service and Maintenance 470 Keysight N6705C Operating and Service Guide ...
Page 471: ...Front Panel Parts Keysight N6705C Operating and Service Guide 471 9 Service and Maintenance ...
Page 476: ......