[SOURce:]ARB:SEQuence:RESet (@<chanlist>)
Resets the Arb sequence to its power-on default setting:
Step = 0
Shape = PULSe
Parameter
Typical Return
(none)
(none)
Returns the number of steps in the Arb sequence: ARB:SEQ:LENG? (@1)
[SOURce:]ARB:SEQuence:STEP:COUNt <value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:COUNt? [MIN|MAX,] <step#>, (@<chanlist>)
Sets the number of times that the sequence step is repeated. The repeat count range is 1 through >16
million. The maximum number of steps that can be programmed is 100. This setting is only valid if the
[SOURce:]ARB:SEQuence:STEP:PACing is TRIGgered
Parameter
Typical Return
Count: 1 - 16,777,216 | MIN | MAX | INFinity, *RST 1
<step count>
Step: 0 - 99, *RST 0
(none)
Sets the a repeat count of 10 for step 1: ARB:SEQ:STEP:COUN 10, 1, (@1)
l
Use the INFinity parameter to repeat the sequence step continuously. If MAX, or a value greater
than 16,777,216 is programmed, the step will also repeat continuously.
l
Use
to stop the sequence at any time. When the sequence is aborted, the out-
put returns to the settings that were in effect before the sequence started.
l
If a count of 9.9E37 is returned, it means that the sequence step is set to repeat continuously.
[SOURce:]ARB:SEQuence:STEP:CURRent
[SOURce:]ARB:SEQuence:STEP:VOLTage
Program waveform steps within an Arb sequence.
All of the [SOURce:]ARB:CURRent and [SOURce:]ARB:VOLTage commands that set waveform
parameters have corresponding commands that set those same parameters for waveform steps
within an Arb sequence. They follow the pattern:
[SOURce:]ARB:CURRent:EXPonential:END <level>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:CURRent:EXPonential:END <level>, <step#>, (@<chanlist>)
where items in bold are added to the sequence step version of the single Arb command. The <step#>
parameter indicates the step number within the Arb sequence. The following Arb functions use the
ARB:SEQuence <step#> format:
,
,
,
, and
.
7 SCPI Programming Reference
264
Keysight N6705C Operating and Service Guide
Summary of Contents for N6705C
Page 76: ......
Page 353: ...Status Diagram Keysight N6705C Operating and Service Guide 353 7 SCPI Programming Reference ...
Page 376: ...8 Verification and Calibration 376 Keysight N6705C Operating and Service Guide ...
Page 389: ...Keysight N6705C Operating and Service Guide 389 8 Verification and Calibration ...
Page 434: ......
Page 438: ...9 Service and Maintenance 438 Keysight N6705C Operating and Service Guide ...
Page 460: ...9 Service and Maintenance 460 Keysight N6705C Operating and Service Guide ...
Page 461: ...Keysight N6705C Operating and Service Guide 461 9 Service and Maintenance ...
Page 462: ...9 Service and Maintenance 462 Keysight N6705C Operating and Service Guide ...
Page 463: ...Keysight N6705C Operating and Service Guide 463 9 Service and Maintenance ...
Page 470: ...Mezzanine Parts 9 Service and Maintenance 470 Keysight N6705C Operating and Service Guide ...
Page 471: ...Front Panel Parts Keysight N6705C Operating and Service Guide 471 9 Service and Maintenance ...
Page 476: ......