Trigger Source
Description
Voltage <1-4> level
Current <1-4> level
Triggers the measurement when the voltage or current of the corresponding output passes
through the specified level.
Arb Run/Stop key
Triggers the measurement when the Arb Run/Stop key is pressed.
Output On/Off key
Triggers the measurement when any of the Output On/Off keys are pressed. Also applies to the All
Outputs On/Off key.
BNC Trigger In
Provide a low-true signal to the BNC trigger input connector. The signal must have a minimum
pulse width of 2 microseconds. Selecting BNC Trigger In also enables any digital I/O pins con-
figured as
Remote Command
Send a trigger command over one of the three interfaces (i.e. *TRG).
If a trigger source is grayed out, it is unavailable. This can happen on power modules that cannot
simultaneously display voltage and current. For these power modules, if one of the traces has been
turned on, you cannot use the other trace as the trigger source. Also, current levels are not available
as trigger sources on outputs that have been grouped (paralleled).
Level lets you specify a trigger level if you selected a Voltage level or Current level as the trigger
source.
Slope specifies if the measurement will be triggered on the positive (up-slope) or negative (down-
slope) portion of the waveform.
Mode dropdown lets you select a trigger mode.
Trigger Mode
Description
Auto
Configures the scope to display a single-sweep measurement when a trigger is received, or auto-
matically if a trigger is not received. The scope continues running and waits for another trigger
when the measurement completes.
Single
Configures the scope to display a single-sweep measurement when a trigger is received. The
scope stops running when the measurement completes.
Triggered
Configures the scope to display a single-sweep measurement when a trigger is received. The
scope continues running and waits for the next trigger when the measurement completes.
Scope Ranges
For outputs that have multiple measurement ranges, you can select a lower range for better
measurement resolution. Scope range settings are independent of the Meter View and Data Logger
range settings.
In the Scope Ranges area, select the desired lower measurement range from the Voltage or Current
dropdown menus.
Keysight N6705C Operating and Service Guide
141
4 Using the Measurement Functions
Summary of Contents for N6705C
Page 76: ......
Page 353: ...Status Diagram Keysight N6705C Operating and Service Guide 353 7 SCPI Programming Reference ...
Page 376: ...8 Verification and Calibration 376 Keysight N6705C Operating and Service Guide ...
Page 389: ...Keysight N6705C Operating and Service Guide 389 8 Verification and Calibration ...
Page 434: ......
Page 438: ...9 Service and Maintenance 438 Keysight N6705C Operating and Service Guide ...
Page 460: ...9 Service and Maintenance 460 Keysight N6705C Operating and Service Guide ...
Page 461: ...Keysight N6705C Operating and Service Guide 461 9 Service and Maintenance ...
Page 462: ...9 Service and Maintenance 462 Keysight N6705C Operating and Service Guide ...
Page 463: ...Keysight N6705C Operating and Service Guide 463 9 Service and Maintenance ...
Page 470: ...Mezzanine Parts 9 Service and Maintenance 470 Keysight N6705C Operating and Service Guide ...
Page 471: ...Front Panel Parts Keysight N6705C Operating and Service Guide 471 9 Service and Maintenance ...
Page 476: ......