Parameter
Typical Return
0|OFF|1|ON, *RST OFF
0 or 1
Enable seamless Elog measurement autoranging:
SENS:ELOG:CURR:RANG:AUTO ON, (@1)
SENS:ELOG:VOLT:RANG:AUTO ON, (@1)
l
Autoranging does not include the 10 μA range, which must be selected individually.
SENSe:ELOG:FUNCtion:CURRent 0|OFF|1|ON, (@<chanlist>)
SENSe:ELOG:FUNCtion:CURRent? (@<chanlist>)
SENSe:ELOG:FUNCtion:VOLTage 0|OFF|1|ON, (@<chanlist>)
SENSe:ELOG:FUNCtion:VOLTage? (@<chanlist>)
Enables or disables the Elog current or voltage measurement function.
Parameter
Typical Return
0|OFF|1|ON, *RST current OFF; voltage ON
0 or 1
Enables Elog current measurements: SENS:ELOG:FUNC:CURR ON, (@1)
Enables Elog voltage measurements:SENS:ELOG:FUNC:VOLT ON, (@1)
SENSe:ELOG:FUNCtion:CURRent:MINMax 0|OFF|1|ON, (@<chanlist>)
SENSe:ELOG:FUNCtion:CURRent:MINMax? (@<chanlist>)
SENSe:ELOG:FUNCtion:VOLTage:MINMax 0|OFF|1|ON, (@<chanlist>)
SENSe:ELOG:FUNCtion:VOLTage:MINMax? (@<chanlist>)
Enables or disables logging of the minimum and maximum current or voltage values.
Parameter
Typical Return
0|OFF|1|ON, *RST OFF
0 or 1
Enables MIN/MAX Elog current values: SENS:ELOG:FUNC:CURR:MINM ON, (@1)
Enables MIN/MAX Elog voltage values:SENS:ELOG:FUNC:VOLT:MINM ON, (@1)
SENSe:ELOG:PERiod <
value
>, (@<chanlist>)
SENSe:ELOG:PERiod? [MIN|MAX,] (@<chanlist>)
Sets the integration time of an Elog measurement.
Although the absolute minimum logging period is 102.4 microseconds, the actual minimum varies as a
function of the number of readings that are being logged (see
).
Parameter
Typical Return
0.0001024 to 60|MIN|MAX, *RST 0.1
<period>
Specifies an Elog period of 0.01 seconds: SENS:ELOG:PER 0.01, (@1)
Keysight N6705C Operating and Service Guide
321
7 SCPI Programming Reference
Summary of Contents for N6705C
Page 76: ......
Page 353: ...Status Diagram Keysight N6705C Operating and Service Guide 353 7 SCPI Programming Reference ...
Page 376: ...8 Verification and Calibration 376 Keysight N6705C Operating and Service Guide ...
Page 389: ...Keysight N6705C Operating and Service Guide 389 8 Verification and Calibration ...
Page 434: ......
Page 438: ...9 Service and Maintenance 438 Keysight N6705C Operating and Service Guide ...
Page 460: ...9 Service and Maintenance 460 Keysight N6705C Operating and Service Guide ...
Page 461: ...Keysight N6705C Operating and Service Guide 461 9 Service and Maintenance ...
Page 462: ...9 Service and Maintenance 462 Keysight N6705C Operating and Service Guide ...
Page 463: ...Keysight N6705C Operating and Service Guide 463 9 Service and Maintenance ...
Page 470: ...Mezzanine Parts 9 Service and Maintenance 470 Keysight N6705C Operating and Service Guide ...
Page 471: ...Front Panel Parts Keysight N6705C Operating and Service Guide 471 9 Service and Maintenance ...
Page 476: ......