4-Wire Sense Connections
Over-Voltage Protection Considerations
Wiring
The power analyzer includes built-in relays that connect or disconnect the ± sense terminals from
their corresponding ± output terminals. As shipped, the sense terminals are internally connected to
the output terminals. This is referred to as Local sensing.
4-wire, or remote sensing improves the voltage regulation at the load by monitoring the voltage at the
load rather than at the output terminals. This automatically compensates for the voltage drop in the
load leads, which is especially useful for CV operation with load impedances that vary or have
significant lead resistance. Because remote sensing is independent of other power analyzer functions,
it can be used regardless of how the instrument is programmed. Remote sensing has no effect during
CC operation
The following figures illustrate load connections using local sensing (A), and 4-wire remote sensing
(B). When the 4-wire indicator above the sense terminals is on, it indicates that the sense terminals
must be connected to the load.
1. Twist leads
2. 4-wire disabled
(indicator is off)
3. 4-wire enabled
(indicator is on)
Connect the load to the output terminals using separate connecting wires. Keep the wire-pair as short
as possible and twist or bundle it to reduce lead inductance and noise pickup. Keep the load leads
under 14.7 meters (50 feet) per lead because of inductance effects.
2 Installation
62
Keysight N6705C Operating and Service Guide
Summary of Contents for N6705C
Page 76: ......
Page 353: ...Status Diagram Keysight N6705C Operating and Service Guide 353 7 SCPI Programming Reference ...
Page 376: ...8 Verification and Calibration 376 Keysight N6705C Operating and Service Guide ...
Page 389: ...Keysight N6705C Operating and Service Guide 389 8 Verification and Calibration ...
Page 434: ......
Page 438: ...9 Service and Maintenance 438 Keysight N6705C Operating and Service Guide ...
Page 460: ...9 Service and Maintenance 460 Keysight N6705C Operating and Service Guide ...
Page 461: ...Keysight N6705C Operating and Service Guide 461 9 Service and Maintenance ...
Page 462: ...9 Service and Maintenance 462 Keysight N6705C Operating and Service Guide ...
Page 463: ...Keysight N6705C Operating and Service Guide 463 9 Service and Maintenance ...
Page 470: ...Mezzanine Parts 9 Service and Maintenance 470 Keysight N6705C Operating and Service Guide ...
Page 471: ...Front Panel Parts Keysight N6705C Operating and Service Guide 471 9 Service and Maintenance ...
Page 476: ......