Front Panel Menu Reference
SCPI Command
From the Calibration menu, select Miscellaneous, Downprogrammer. Select
Start. Then select Close.
Enter a Calibration Date, Save, and Log Out
Front Panel Menu Reference
SCPI Command
From the Calibration menu, select Date. Enter the date, then select Insert Date.
Then select Close.
From the Calibration menu, select Save. Select Save Calibration for all Outputs.
Then select Close.
From the Administrative Tools menu, select Administrator Login/Logout to log
out of the Admin menu.
To enter a calibration date:
To save the calibration data:
To exit calibration mode:
8 Verification and Calibration
398
Keysight N6705C Operating and Service Guide
Summary of Contents for N6705C
Page 76: ......
Page 353: ...Status Diagram Keysight N6705C Operating and Service Guide 353 7 SCPI Programming Reference ...
Page 376: ...8 Verification and Calibration 376 Keysight N6705C Operating and Service Guide ...
Page 389: ...Keysight N6705C Operating and Service Guide 389 8 Verification and Calibration ...
Page 434: ......
Page 438: ...9 Service and Maintenance 438 Keysight N6705C Operating and Service Guide ...
Page 460: ...9 Service and Maintenance 460 Keysight N6705C Operating and Service Guide ...
Page 461: ...Keysight N6705C Operating and Service Guide 461 9 Service and Maintenance ...
Page 462: ...9 Service and Maintenance 462 Keysight N6705C Operating and Service Guide ...
Page 463: ...Keysight N6705C Operating and Service Guide 463 9 Service and Maintenance ...
Page 470: ...Mezzanine Parts 9 Service and Maintenance 470 Keysight N6705C Operating and Service Guide ...
Page 471: ...Front Panel Parts Keysight N6705C Operating and Service Guide 471 9 Service and Maintenance ...
Page 476: ......