[SOURce:]LIST:STEP ONCE|AUTO, (@<chanlist>)
[SOURce:]LIST:STEP? (@<chanlist>)
Specifies how the list responds to triggers.
ONCE - the output remains at the present step until a trigger advances it to the next step. Triggers
that arrive during the dwell time are ignored.
AUTO - the output automatically advances to each step, after the receipt of an initial starting trigger.
Steps are paced by the dwell list. As each dwell time elapses, the next step is immediately output.
Parameter
Typical Return
ONCE|AUTO, *RST AUTO
ONCE or AUTO
Specifies the list steps to be paced by trigger signals: LIST:STEP ONCE, (@1)
[SOURce:]LIST:TERMinate:LAST 0|OFF|1|ON, (@<chanlist>)
[SOURce:]LIST:TERMinate:LAST? (@<chanlist>)
Determines the output value when the list terminates. When ON (1), the output remains at the last list
step. The value of the last list step becomes the IMMediate value. When OFF (0), and also when the
list is aborted, the output returns to the settings that were in effect before the list started.
Parameter
Typical Return
0|OFF|1|ON , *RST OFF
0 or 1
Terminate with the output at the last step value: LIST:TERM:LAST ON, (@1)
l
The query command returns 0 if the output returns to the settings that were in effect before the list
started, and 1 if the output voltage or current remains at the value of the last list step.
[SOURce:]LIST:TOUTput:BOSTep[:DATA] 0|OFF|1|ON {,0|OFF|1|ON}, (@<chanlist>)
[SOURce:]LIST:TOUTput:BOSTep[:DATA]? (@<chanlist>)
[SOURce:]LIST:TOUTput:EOSTep[:DATA] 0|OFF|1|ON {,0|OFF|1|ON}, (@<chanlist>)
[SOURce:]LIST:TOUTput:EOSTep[:DATA]? (@<chanlist>)
Specifies which list steps generate a trigger signal at the beginning of step (BOSTep) or end of step
(EOSTep). A comma-delimited list of up to 512 steps may be programmed.
A trigger is only generated when the state is set to ON. The trigger signal can be used as a trigger
source for measurements and transients of other units, and for digital port pins configured as trigger
outputs.
7 SCPI Programming Reference
298
Keysight N6705C Operating and Service Guide
Summary of Contents for N6705C
Page 76: ......
Page 353: ...Status Diagram Keysight N6705C Operating and Service Guide 353 7 SCPI Programming Reference ...
Page 376: ...8 Verification and Calibration 376 Keysight N6705C Operating and Service Guide ...
Page 389: ...Keysight N6705C Operating and Service Guide 389 8 Verification and Calibration ...
Page 434: ......
Page 438: ...9 Service and Maintenance 438 Keysight N6705C Operating and Service Guide ...
Page 460: ...9 Service and Maintenance 460 Keysight N6705C Operating and Service Guide ...
Page 461: ...Keysight N6705C Operating and Service Guide 461 9 Service and Maintenance ...
Page 462: ...9 Service and Maintenance 462 Keysight N6705C Operating and Service Guide ...
Page 463: ...Keysight N6705C Operating and Service Guide 463 9 Service and Maintenance ...
Page 470: ...Mezzanine Parts 9 Service and Maintenance 470 Keysight N6705C Operating and Service Guide ...
Page 471: ...Front Panel Parts Keysight N6705C Operating and Service Guide 471 9 Service and Maintenance ...
Page 476: ......