181
Programming Commands
CALCulate3 Subsystem
DELTa:REFerence[:WAVelength]
Selects the reference laser line for DELTa calculations.
Syntax
:CALCulate3:DELTa:REFerence[:WAVelength]{?| {<real> | MINimum | MAXimum}}
<real>
is a wavelength value that is within the following limits:
Constant
Description
MINimum
700.0 nm
MAXimum
1650.0 nm
Attribute Summary
Preset State: 700 nm (428.6 THz)
*RST State: 700 nm (428.6 THz) laser line
SCPI Compliance: instrument specific
Description
The reference will be the laser line at the wavelength closest to the wavelength
entered. Subsequent measurements will use the wavelength closest to the reference
wavelength used for the previous measurement.
The query returns the current wavelength of the reference laser line. The default
units for the
<real>
parameter are meters.